Arterial compliance is a main variable determining the effectiveness of intra-aortic balloon counterpulsation: quantitative data from an in vitro study

被引:31
作者
Papaioannou, TG [1 ]
Mathioulakis, DS
Nanas, JN
Tsangaris, SG
Stamatelopoulos, SF
Moulopoulos, SD
机构
[1] Natl Univ Athens, Sch Med, Dept Clin Therapeut, Athens, Greece
[2] Natl Tech Univ Athens, Dept Mech Engn, GR-10682 Athens, Greece
关键词
mock circulation; counterpulsation; assisted circulation;
D O I
10.1016/S1350-4533(02)00013-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Quantitative data concerning the effect of arterial compliance (AC) on the effectiveness of intra-aortic balloon counterpulsation (IABC) are lacking. The main objective of this study was to investigate the relationship between AC and IABC performance. For this purpose we constructed a Windkessel, lumped-element, hydraulic model of the systemic circulation. The model consisted of a left ventricular assist device (LVAD), a compliance chamber, a peripheral resistor and two open reservoirs. Two Datascope Driving systems were used to operate the LVAD and intra-aortic balloon. We studied the effect of arterial compliance on the effectiveness of IABC at different levels of mean pressure (55, 75 and 95 mmHg) and heart rates (80, 100, 120 bpm). Three indices were used to evaluate IABC performance: the reduction of systolic and end-diastolic "arterial" pressure and the augmentation of diastolic pressure, induced by the IABC. A 22% decrease in AC (1.8-1.4 ml/mmHg) lead to a 30-40% increase in the indices of IABC performance, independently from pressure. In conclusion, arterial compliance significantly affects IABC efficacy and it could be considered as a further clinical criterion to decide IABC application. (C) 2002 IPEM. Published by Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:279 / 284
页数:6
相关论文
共 19 条
[1]   Wave propagation in coupled left ventricle-arterial system - Implications for aortic pressure [J].
Berger, DS ;
Robinson, KA ;
Shroff, SG .
HYPERTENSION, 1996, 27 (05) :1079-1089
[2]   DEVELOPMENT OF MOCK CIRCULATION MODELS FOR THE ASSESSMENT OF COUNTERPULSATION SYSTEMS [J].
BOWLES, CT ;
NISHIMURA, K ;
SHAH, SS ;
CUMMING, DVE ;
PEPPER, JR ;
YACOUB, MH ;
CLARK, C ;
PATTISON, CW .
CARDIOVASCULAR RESEARCH, 1991, 25 (11) :901-908
[3]   Use of radial artery applanation tonometry and a generalized transfer function to determine aortic pressure augmentation in subjects with treated hypertension [J].
Cameron, JD ;
McGrath, BP ;
Dart, AM .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 32 (05) :1214-1220
[4]  
Charitos CE, 1998, J CARDIOVASC SURG, V39, P625
[5]   THEORETICAL CONSIDERATIONS REGARDING THE OPTIMIZATION OF CARDIAC ASSISTANCE BY INTRAAORTIC BALLOON PUMPING [J].
JARON, D ;
MOORE, TW ;
HE, P .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (03) :177-186
[6]   INITIAL CLINICAL EXPERIENCE WITH INTRAAORTIC BALLOON PUMPING IN CARDIOGENIC SHOCK [J].
KANTROWITZ, A ;
TJONNELAND, S ;
FREED, PS ;
PHILLIPS, SJ ;
BUTNER, AN ;
SHERMAN, JL .
JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1968, 203 (02) :113-+
[7]   INFLUENCE OF VASCULAR PARAMETERS ON THE EFFECTIVENESS OF INTRA-AORTIC BALLOON PUMPING - A MODEL STUDY [J].
KAO, C ;
OHLEY, WJ .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1982, 20 (05) :529-538
[8]  
Lin C Y, 1972, Ann Thorac Surg, V13, P468
[9]  
MAHON T, 1971, J BIOMECH, V4, P335
[10]   NONINVASIVE METHOD FOR DETERMINATION OF ARTERIAL COMPLIANCE USING DOPPLER-ECHOCARDIOGRAPHY AND SUBCLAVIAN PULSE TRACINGS - VALIDATION AND CLINICAL-APPLICATION OF A PHYSIOLOGICAL MODEL OF THE CIRCULATION [J].
MARCUS, RH ;
KORCARZ, C ;
MCCRAY, G ;
NEUMANN, A ;
MURPHY, M ;
BOROW, K ;
WEINERT, L ;
BEDNARZ, J ;
GRETLER, DD ;
SPENCER, KT ;
SARELI, P ;
LANG, RM .
CIRCULATION, 1994, 89 (06) :2688-2699