Pump-induced haemolysis: a comparison of short-term ventricular assist devices

被引:42
作者
Bennett, M [1 ]
Horton, S [1 ]
Thuys, C [1 ]
Augustin, S [1 ]
Rosenberg, M [1 ]
Brizard, C [1 ]
机构
[1] Royal Childrens Hosp, Cardiac Surg Unit, Parkville, Vic 3052, Australia
来源
PERFUSION-UK | 2004年 / 19卷 / 02期
关键词
D O I
10.1191/0267659104pf729oa
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Centrifugal pumps are superior to roller pumps for extended support durations in terms of pump-induced haemolysis. In this study, we evaluated the commonly used Biomedicus BP 50 and compared it with the Jostra Rotaflow and a standard roller pump in an in vitro test circuit. Each circuit was run for a six-day period and repeated five times. Plasma haemoglobin values showed the roller pump to become more haemolytic than the Biomedicus (p = 0.022) and the Rotaflow. A statistically significant difference between the Biomedicus and the Rotaflow was observed on day six of the trial (p = 0.016), with the Rotaflow showing lower levels of haemolysis than the Biomedicus. These results support the use of the new generation centrifugal pump, the Rotaflow, as a suitable device for short-term ventricular assist.
引用
收藏
页码:107 / 111
页数:5
相关论文
共 14 条
[1]   RECENT STUDIES OF THE CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER [J].
AKAMATSU, T ;
TSUKIYA, T ;
NISHIMURA, K ;
PARK, CH ;
NAKAZEKI, T .
ARTIFICIAL ORGANS, 1995, 19 (07) :631-634
[2]  
[Anonymous], JECT
[3]  
BURTIS CA, 1994, TIETZ TXB CLIN CHEM, P2022
[4]   Centrifugal pumps: Description of devices and surgical techniques [J].
Curtis, JJ ;
Walls, JT ;
Wagner-Mann, CC ;
Schmaltz, RA ;
Demmy, TL ;
McKenney, CA ;
Mann, FA .
ANNALS OF THORACIC SURGERY, 1999, 68 (02) :666-671
[5]  
*ELSO, 2003, ECLS REG REP INT SUM
[6]  
HORTON A M, 1992, Perfusion (London), V7, P103, DOI 10.1177/026765919200700204
[7]  
HORTON MA, 1993, J BONE MINER RES, V8, P239
[8]  
Kawahito K, 1997, ARTIF ORGANS, V21, P323
[9]   The role of the centrifugal pump in hemolysis during neonatal extracorporeal support [J].
McDonald, JV ;
Green, TP ;
Steinhorn, RH .
ASAIO JOURNAL, 1997, 43 (01) :35-38
[10]   SEAL-LESS CENTRIFUGAL BLOOD PUMP WITH MAGNETICALLY SUSPENDED ROTOR - ROT-A-FLOT [J].
MENDLER, N ;
PODECHTL, F ;
FEIL, G ;
HILTMANN, P ;
SEBENING, F .
ARTIFICIAL ORGANS, 1995, 19 (07) :620-624