SEAL-LESS CENTRIFUGAL BLOOD PUMP WITH MAGNETICALLY SUSPENDED ROTOR - ROT-A-FLOT

被引:47
作者
MENDLER, N
PODECHTL, F
FEIL, G
HILTMANN, P
SEBENING, F
机构
[1] Department of Cardiovascular Surgery, German Heart Center, Munich
关键词
CENTRIFUGAL BLOOD PUMP; SEAL-LESS PUMP; MAGNETIC SUSPENSION; EFFICIENCY OF BLOOD PUMP; COMPARISON OF BLOOD PUMPS;
D O I
10.1111/j.1525-1594.1995.tb02391.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Limitations of current centrifugal blood pumps are related to heat generation of bearings and leakage of seals, to dead water zones, and to poor efficiency. A new concept is proposed in this paper to ameliorate these problems based on a miniaturized magnetic drive, and a prototype is introduced. The pump rotor is suspended and driven by a radial permanent magnetic field that stabilizes the impeller in 4 of the 6 spatial degrees of freedom and allows it to be top-spun on a single blood-flushed pivot bearing with minimal load and friction. A shrouded impeller with an open center and 4 logarithmically curved channels is run inside a cone-and-plate-type housing with a spiral volute chamber. In vitro testing was performed comparing this design with the BioMedicus, St. Jude, and Sarns pumps. The prototype is demonstrated to have the smallest internal volume (35 ml), surface (190 gcm), and passage time (0.5 s at 4 L/min), as well as the highest hydraulic efficiency (up to 0.4) of all devices studied.
引用
收藏
页码:620 / 624
页数:5
相关论文
共 14 条
[1]  
AKAMATSU T, 1993, HEART REPLACEMENT AR, P305
[2]   A FLOW VISUALIZATION STUDY OF THE NCVC CENTRIFUGAL BLOOD PUMP [J].
ARAKI, K ;
TAENAKA, Y ;
MASUZAWA, T ;
TATSUMI, E ;
WAKISAKA, Y ;
WATARI, M ;
NAKATANI, T ;
AKAGI, H ;
BABA, Y ;
ANAI, H ;
PARK, YH ;
EYA, K ;
TODA, K ;
TAKANO, H .
ARTIFICIAL ORGANS, 1994, 18 (09) :669-672
[3]  
BLOOMFIELD DA, 1974, DYE CURVES THEORY PR
[4]   FREQUENCY OF SEAL DISRUPTION WITH THE SARNS CENTRIFUGAL PUMP IN POSTCARDIOTOMY CIRCULATORY ASSIST [J].
CURTIS, JJ ;
BOLEY, TM ;
WALLS, JT ;
DEMMY, TL ;
SCHMALTZ, RA .
ARTIFICIAL ORGANS, 1994, 18 (03) :235-237
[5]  
ENGELHARDT H, 1988, P INT WORKSH ROT BLO, P76
[6]  
KIJIMA T, 1993, HEART REPLACEMENT, P309
[7]  
MENDLER N, 1994, Patent No. 4430853
[8]  
OHARA Y, 1993, NONPULSATILE PERFUSI, P38
[9]  
Olsen D B, 1985, Trans Am Soc Artif Intern Organs, V31, P395
[10]  
RHEE K, 1991, P INT WORKSH ROT BLO, P9