The development of a national standardized approach for the enumeration of residual leucocytes in blood components

被引:11
作者
Cardigan, R
Phipps, A
Seghatchian, J
Bashir, S
Aynsley, S
Beckman, N
Barnett, D
Reilly, JT
Williamson, LM
机构
[1] Natl Blood Serv, Brentwood CM15 8DP, Essex, England
[2] Natl Blood Serv, Bristol, Avon, England
[3] Natl Blood Serv, Newcastle, NSW, Australia
[4] Royal Hallamshire Hosp, Dept Haematol, UK NEQAS Leucocyte Immunophenotyping, Sheffield S10 2JF, S Yorkshire, England
[5] Univ Cambridge, Dept Haematol, Cambridge CB2 1TN, England
关键词
count; flow cytometry; leucocyte depletion; standardization; statistical process control;
D O I
10.1046/j.1423-0410.2002.00194.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Objectives The UK Blood Transfusion Services implemented universal leucocyte depletion of the blood supply in November 1999. To provide statistical process monitoring of these processes, automated methods were introduced to count residual leucocytes (white blood cells) in blood components. Materials and Methods Initially in the National Blood Service (NBS) England, protocols were standardized on the use of LeucoCount reagents with either Becton-Dickinson or Beckman Coulter flow cytometers. Results Standardization of protocols resulted in a decreased intersite variability of red cell samples (from 36% to 9% at a level of 11 and 10 cells/mul, respectively), and 100% of sites (n = 11) fulfilled the validation criteria. However, we also evaluated the use of alternative reagents with the result that reagents from either Becton-Dickinson or Beckman Coulter, used on either a Becton-Dickinson or Beckman Coulter flow cytometer, passed our validation criteria. Conclusions It is critical to include samples from filtered products containing white blood cells in validations of leucocyte enumeration methodology, as results may differ between methods using these samples but not using spiked or fixed material. Standardized gating strategies and optimization methods for flow cytometers are critical for obtaining equivalent results with different reagents and instruments.
引用
收藏
页码:100 / 109
页数:10
相关论文
共 16 条
[1]   Flow cytometric determination of residual leucocytes in filter-depleted blood products: An evaluation of Becton-Dickinson's LeucoCOUNT systems [J].
Barclay, R ;
Walker, B ;
Allan, R ;
Reid, C ;
Duffin, E ;
Kane, E ;
Turner, M .
TRANSFUSION SCIENCE, 1998, 19 (04) :399-403
[2]   Low level leucocyte counting: a critical variable in the validation of leucodepleted blood transfusion components as highlighted by an external quality assessment study [J].
Barnett, D ;
Goodfellow, K ;
Ginnever, J ;
Granger, V ;
Whitby, L ;
Reilly, JT .
CLINICAL AND LABORATORY HAEMATOLOGY, 2001, 23 (01) :43-51
[3]  
BARNETT D, 1998, Patent No. 2279653
[4]   RELIABLE AND VERY SENSITIVE FLOW-CYTOMETRIC METHOD FOR COUNTING LOW LEUKOCYTE NUMBERS IN PLATELET CONCENTRATES [J].
DENEYS, V ;
MAZZON, AM ;
ROBERT, A ;
DUVILLIER, H ;
DEBRUYERE, M .
VOX SANGUINIS, 1994, 67 (02) :172-177
[5]   Practical guidelines for process validation and process control of white cell-reduced blood components: Report of the biomedical excellence for safer transfusion (BEST) working party of the international society of blood transfusion (ISBT) [J].
Dumont, LJ ;
Dzik, WH ;
Rebulla, P ;
Brandwein, H .
TRANSFUSION, 1996, 36 (01) :11-20
[6]   A multicenter study evaluating three methods for counting residual WBCs in WBC-reduced blood components: Nageotte hemocytometry, flow cytometry, and microfluorometry [J].
Dzik, S ;
Moroff, G ;
Dumont, L .
TRANSFUSION, 2000, 40 (05) :513-520
[7]  
Dzik WH, 2000, VOX SANG, V78, P223
[8]  
Gómez-Arbonés X, 1999, VOX SANG, V76, P64
[9]  
GOODFELLOW KJ, 2002, IN PRESS TRANSFUSION
[10]   VALIDATION OF A SIMPLE METHOD TO COUNT VERY LOW WHITE CELL CONCENTRATIONS IN FILTERED RED-CELLS OR PLATELETS [J].
MASSE, M ;
NAEGELEN, C ;
PELLEGRINI, N ;
SEGIER, JM ;
MARPAUX, N ;
BEAUJEAN, F .
TRANSFUSION, 1992, 32 (06) :565-571