Real-time contrast echo assessment of myocardial perfusion at low emission power:: First experimental and clinical results using power pulse inversion imaging

被引:130
作者
Tiemann, K [1 ]
Lohmeier, S [1 ]
Kuntz, S [1 ]
Köster, J [1 ]
Pohl, C [1 ]
Burns, P [1 ]
Porter, TR [1 ]
Nanda, NC [1 ]
Lüderitz, B [1 ]
Becher, H [1 ]
机构
[1] Univ Bonn, Dept Cardiol, D-53105 Bonn, Germany
来源
ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES | 1999年 / 16卷 / 08期
关键词
power pulse inversion imaging; contrast echocardiography; myocardial perfusion; low emission power;
D O I
10.1111/j.1540-8175.1999.tb00132.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Power pulse inversion (PPI) has been, developed for echocontrast specific imaging in order to reduce destruction of microbubbles. The purpose of this study was to evaluate PPI for real-time contrast echocardiography. Therefore, in, vitro studies in a physiological flow-phantom and clinical examinations in, patients with coronary artery disease were performed. The in vitro results of this study indicate that PPI allows real-time imaging at low emission power and is almost nondestructive to contrast microbubbles of Definity. At this low emission power a strong linear relationship between the dosage of the contrast agent and the resulting PPI signal intensity was found (R = 0.998, p < 0.001). In the clinical examinations real-time imaging using low mechanical index PPI resulted in strong myocardial signals and a complete filling of the cavities indicating absence of bubble destruction. Most striking was the ability of PPI to display myocardial thickening and wall motion simultaneously with the assessment of myocardial contrast replenishment following ultrasound induced bubble destruction by high power frames. We conclude that PPI allows nondestructive contrast imaging both in experimental and clinical settings. Therefore, real-time imaging of myocardial perfusion, and real-time assessment of contrast replenishment following ultrasound induced destruction of microbubbles is feasible. Moreover, PPI allows simultaneous assessment of perfusion and myocardial function.
引用
收藏
页码:799 / 809
页数:11
相关论文
共 17 条
[1]   Measurements of harmonic generation in a focused finite-amplitude sound beam [J].
Averkiou, MA ;
Hamilton, MF .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1995, 98 (06) :3439-3442
[2]   Improvement in endocardial border delineation using tissue harmonic imaging [J].
Becher, H ;
Tiemann, K ;
Schlosser, T ;
Pohl, C ;
Nanda, NC ;
Averkiou, MA ;
Powers, J ;
Lüderitz, B .
ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 1998, 15 (05) :511-517
[3]  
Burns PN, 1996, CLIN RADIOL, V51, P50
[4]   TOPOLOGY AND DIMENSIONS OF PIG CORONARY CAPILLARY NETWORK [J].
KASSAB, GS ;
FUNG, YCB .
AMERICAN JOURNAL OF PHYSIOLOGY, 1994, 267 (01) :H319-H325
[5]   Analysis of pig's coronary arterial blood flow with detailed anatomical data [J].
Kassab, GS ;
Berkley, J ;
Fung, YCB .
ANNALS OF BIOMEDICAL ENGINEERING, 1997, 25 (01) :204-217
[6]  
Kaul S, 1997, CIRCULATION, V96, P719
[7]  
MARCUS M, 1999, CORONARY CIRCULATION, P3
[8]   An investigation of the relationship between ultrasound echo enhancement and Doppler frequency shift using a pulsatile arterial flow phantom [J].
Petrick, J ;
Zomack, M ;
Schlief, R .
INVESTIGATIVE RADIOLOGY, 1997, 32 (04) :225-235
[9]  
Porter T R, 1996, J Am Soc Echocardiogr, V9, P599
[10]   Real-time visualization of myocardial perfusion and wall thickening in human beings with intravenous ultrasonographic contrast and accelerated intermittent harmonic imaging [J].
Porter, TR ;
Li, SP ;
Jiang, L ;
Grayburn, P ;
Deligonul, U .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 1999, 12 (04) :266-271