摘要
Of all available reconstruction methods, statistical iterative reconstruction algorithms appear particularly promising since they enable accurate physical noise modeling. The newly developed compressive sampling/compressed sensing (CS) algorithm has shown the potential to accurately reconstruct images from highly undersampled data. The CS algorithm can be implemented in the statistical reconstruction framework as well. In this study, we compared the performance of two standard statistical reconstruction algorithms (penalized weighted least squares and q-GGMRF) to the CS algorithm. In assessing the image quality using these iterative reconstructions, it is critical to utilize realistic background anatomy as the reconstruction results are object dependent. A cadaver head was scanned on a Varian Trilogy system at different dose levels. Several figures of merit including the relative root mean square error and a quality factor which accounts for the noise performance and the spatial resolution were introduced to objectively evaluate reconstruction performance. A comparison is presented between the three algorithms for a constant undersampling factor comparing different algorithms at several dose levels. To facilitate this comparison, the original CS method was formulated in the framework of the statistical image reconstruction algorithms. Important conclusions of the measurements from our studies are that (1) for realistic neuro-anatomy, over 100 projections are required to avoid streak artifacts in the reconstructed images even with CS reconstruction, (2) regardless of the algorithm employed, it is beneficial to distribute the total dose to more views as long as each view remains quantum noise limited and (3) the total variationbased CS method is not appropriate for very low dose levels because while it can mitigate streaking artifacts, the images exhibit patchy behavior, which is potentially harmful for medical diagnosis.
引用
收藏
页码:5781 / 5804
页数:24
相关论文
共 71 条
[1]
SIMULTANEOUS ALGEBRAIC RECONSTRUCTION TECHNIQUE (SART) - A SUPERIOR IMPLEMENTATION OF THE ART ALGORITHM [J].
ANDERSEN, AH ;
KAK, AC .
ULTRASONIC IMAGING, 1984, 6 (01) :81-94
[2]
Ordered subset reconstruction for x-ray CT [J].
Beekman, FJ ;
Kamphuis, C .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (07) :1835-1844
[3]
Undersampled radial MRI with multiple coils. Iterative image reconstruction using a total variation constraint [J].
Block, Kai Tobias ;
Uecker, Martin ;
Frahm, Jens .
MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (06) :1086-1098
[4]
A unified approach to statistical tomography using coordinate descent optimization [J].
Bouman, CA ;
Sauer, K .
IEEE TRANSACTIONS ON IMAGE PROCESSING, 1996, 5 (03) :480-492
[5]
A generalized Gaussian image model for edge-preserving MAP estimation [J].
Bournan, Charles ;
Sauer, Ken .
IEEE TRANSACTIONS ON IMAGE PROCESSING, 1993, 2 (03) :296-310
[6]
Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[7]
Robust uncertainty principles:: Exact signal reconstruction from highly incomplete frequency information [J].
Candès, EJ ;
Romberg, J ;
Tao, T .
IEEE TRANSACTIONS ON INFORMATION THEORY, 2006, 52 (02) :489-509
[8]
CHEN GH, 2008, P SPIE, V6856
[9]
Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic CT images from highly undersampled projection data sets [J].
Chen, Guang-Hong ;
Tang, Jie ;
Leng, Shuai .
MEDICAL PHYSICS, 2008, 35 (02) :660-663
[10]
Temporal resolution improvement using PICCS in MDCT cardiac imaging [J].
Chen, Guang-Hong ;
Tang, Jie ;
Hsieh, Jiang .
MEDICAL PHYSICS, 2009, 36 (06) :2130-2135