Bovine type I collagen as an endovascular stent-graft material: Biocompatibility study in rabbits

被引:23
作者
Cloft, HJ
Kallmes, DF
Lin, HB
Li, ST
Marx, WF
Hudson, SB
Helm, GA
Lopes, MB
McGraw, JK
Dion, JE
Jensen, ME
机构
[1] Univ Virginia, Hlth Sci Ctr, Dept Radiol, Charlottesville, VA 22908 USA
[2] Univ Virginia, Hlth Sci Ctr, Dept Neurol Surg, Charlottesville, VA 22908 USA
[3] Univ Virginia, Hlth Sci Ctr, Dept Pathol, Charlottesville, VA 22908 USA
[4] ReGen, Franklin Lakes, NJ USA
关键词
animals; aorta; grafts and prostheses; aortography; interventional procedures; experimental studies; stents and prostheses;
D O I
10.1148/radiology.214.2.r00fe21557
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To study the biocompatibility of a bovine type I collagen preparation as a material for small-vessel stent grafts in rabbits. MATERIALS AND METHODS: A composite nitinol-collagen endovascular stent-graft with a 4-mm inner diameter was deployed in the abdominal aorta in nine rabbits. Angiography was performed, and the rabbits were sacrificed at 1, 2, and 7 days and at 1 and 3 months. The portion of the aorta containing the stent-graft was excised and was histologically evaluated. RESULTS: All stent-grafts were patent at all time points. On days 1, 2, and 7 after implantation, scattered red and white blood cells adhered to the stent-graft. At 1 month, the stent-graft was endothelialized and was infiltrated with fibroblasts that deposited collagen within the interstices of the implanted collagen material. At 3 months, there was additional collagen deposition within the interstices of the stent-graft that did not narrow the lumen of the stent-grafts. CONCLUSION: Type I collagen as a intravascular stent-graft material is biocompatible for at least 3 months in rabbits. It is rapidly endothelialized and does not cause reactive stenosis. As a versatile and biocompatible polymer, collagen is potentially useful in the construction of endovascular stent-grafts for use in human arteries.
引用
收藏
页码:557 / 562
页数:6
相关论文
共 44 条
[1]   Percutaneous closure of an arteriovenous iliac fistula with a Cragg endoluminal graft [J].
Allgayer, B ;
Theiss, W ;
Naundorf, M .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 166 (03) :673-674
[2]  
BAUGH RF, 1979, CLIN HAEMATOL, V8, P3
[3]   PATENCY RATES OF MINUTE VASCULAR REPLACEMENTS - THE GLUTARALDEHYDE MODIFIED MANDRIL-GROWN CONDUIT [J].
CHRISTIE, B ;
KETHARANATHAN, V ;
PERLOFF, LJ .
JOURNAL OF SURGICAL RESEARCH, 1980, 28 (06) :519-532
[4]  
CHRISTIE BA, 1982, ARCH SURG-CHICAGO, V117, P1290
[5]   TRANSFEMORAL ENDOVASCULAR AORTIC GRAFT PLACEMENT [J].
CHUTER, TAM ;
GREEN, RM ;
OURIEL, K ;
FIORE, WM ;
DEWEESE, JA ;
QUERAL, LA ;
CRIADO ;
NATH, RL ;
BERKOWITZ, HD ;
SUMPIO, BE .
JOURNAL OF VASCULAR SURGERY, 1993, 18 (02) :185-197
[6]  
CHVAPIL M, 1970, MOD TRENDS VASC SURG, V1, P120
[7]  
CLOWES AW, 1985, AM J PATHOL, V118, P43
[8]  
Cragg A H, 1993, J Vasc Interv Radiol, V4, P455, DOI 10.1016/S1051-0443(93)71898-9
[9]   PERCUTANEOUS FEMOROPOPLITEAL GRAFT PLACEMENT [J].
CRAGG, AH ;
DAKE, MD .
RADIOLOGY, 1993, 187 (03) :643-648
[10]   NITINOL INTRAVASCULAR STENT - RESULTS OF PRECLINICAL EVALUATION [J].
CRAGG, AH ;
DEJONG, SC ;
BARNHART, WH ;
LANDAS, SK ;
SMITH, TP .
RADIOLOGY, 1993, 189 (03) :775-778