Comparative evaluation of Kodak EDR2 and XV2 films for verification of intensity modulated radiation therapy

被引:70
作者
Dogan, N [1 ]
Leybovich, LB [1 ]
Sethi, A [1 ]
机构
[1] Loyola Univ, Med Ctr, Dept Radiat Oncol, Maywood, IL 60153 USA
关键词
D O I
10.1088/0031-9155/47/22/314
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Film dosimetry provides a convenient tool to determine dose distributions, especially for verification of IMRT plans. However, the film response to radiation shows a significant dependence on depth, energy and field size. that compromise the accuracy of measurements. Kodak's XV2 film has a low saturation dose (similar to100 cGy) and, consequently, a relatively short region of linear dose-response. The recently introduced Kodak extended range EDR2 film was reported to have a linear dose-response region extending to 500 cGy. This increased dose range may be particularly useful in the verification of IMRT plans. In this work, the dependence of Kodak EDR2 film's response on the depth, field size and energy was evaluated and compared with Kodak XV2 film. Co-60, 6 MV, 10 MV and 18 MV beams were used. Field sizes were 2 x 2, 6 x 6, 10 x 10, 14 x 14,18 x 18 and 24 x 24 cm(2). Doses for XV2 and EDR2 films were 80 cGy and 300 cGy, respectively. Optical density was converted to dose using depth-corrected sensitometric (Hurter and Driffield, or H&D) curves. For each field size, XV2 and EDR2 depth-dose curves were compared with ion chamber depth-dose curves. Both films demonstrated similar (within 1%) field size dependence. The deviation from the ion chamber for both films was small for the fields ranging from 2 x 2 to 10 x 10 cm(2) : less than or equal to2% for 6, 10 and 18 MV beams. No deviation was observed for the Co-60 beam. As the field size increased to 24 x 24 cm 2, the deviation became significant for both films: similar to7.5% for Co-60, similar to5% for 6 MV and 10 MV, and similar to6% for 18 MV. During the verification of IMRT plans, EDR2 film showed a better agreement with the calculated dose distributions than the XV2 film.
引用
收藏
页码:4121 / 4130
页数:10
相关论文
共 16 条
[1]   A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering [J].
Burch, SE ;
Kearfott, KJ ;
Trueblood, JH ;
Sheils, WC ;
Yeo, JI ;
Wang, CKC .
MEDICAL PHYSICS, 1997, 24 (05) :775-783
[2]  
Chungbin S, 2001, MED PHYS, V28, P1208
[3]   Variation of sensitometric curves of radiographic films in high energy photon beams [J].
Danciu, C ;
Proimos, BS ;
Rosenwald, JC ;
Mijnheer, BJ .
MEDICAL PHYSICS, 2001, 28 (06) :966-974
[4]  
Ezzell G A, 2001, J Appl Clin Med Phys, V2, P138, DOI 10.1120/1.1386508
[5]   Intensity modulation with the "step and shoot" technique using a commercial MLC:: A planning study [J].
Keller-Reichenbecher, MA ;
Bortfeld, T ;
Levegrün, S ;
Stein, J ;
Preiser, K ;
Schlegel, W .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (05) :1315-1324
[6]  
Leybovich L, 2002, MED PHYS, V29, P1263
[7]  
ROBERT JL, 1999, MED PHYS, V26, P2144
[8]  
SANG GJ, 2002, MED PHYS, V29, P351
[9]   Perturbation of radiotherapy beams by radiographic film: measurements and Monte Carlo simulations [J].
Suchowerska, N ;
Hoban, P ;
Davison, A ;
Metcalfe, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (07) :1755-1765
[10]   Directional dependence in film dosimetry: radiographic and radiochromic film [J].
Suchowerska, N ;
Hoban, P ;
Butson, M ;
Davison, A ;
Metcalfe, P .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (05) :1391-1397