Characterization of the new NSLS infrared microspectroscopy beamline U10B

被引:9
作者
Carr, GL [1 ]
Merlo, O [1 ]
Munsli, M [1 ]
Springer, S [1 ]
Ho, SC [1 ]
机构
[1] Brookhaven Natl Lab, Natl Synchrotron Light Source, Upton, NY 11973 USA
来源
ACCELERATOR-BASED SOURCES OF INFRARED AND SPECTROSCOPIC APPLICATIONS | 1999年 / 3775卷
关键词
infrared microspectroscopy; infrared synchrotron radiation; infrared beamline; diffraction; spatial resolution;
D O I
10.1117/12.366648
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The first of several new infrared beamlines, built on a modified bending magnet port of the NSLS VUV ring, is now operational for mid-infrared microspectroscopy. The port simultaneously delivers 40 mrad by 40 mrad to two separate beamlines and spectrometer endstations designated U10A and U10B. The latter is equipped with a scanning infrared microspectrometer. The combination of this instrument and high brightness synchrotron radiation makes diffraction-limited microspectroscopy practical. This paper describes the beamline's performance and presents quantitative information on the diffraction-limited resolution.
引用
收藏
页码:22 / 28
页数:7
相关论文
共 8 条
[1]   PERFORMANCE OF AN INFRARED MICROSPECTROMETER AT THE NSLS [J].
CARR, GL ;
REFFNER, JA ;
WILLIAMS, GP .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02) :1490-1492
[2]  
Carr GL, 1997, P SOC PHOTO-OPT INS, V3153, P51, DOI 10.1117/12.290262
[3]  
Carr GL, 1996, SEMICONDUCTOR CHARACTERIZATION, P418
[4]   INFRARED SYNCHROTRON RADIATION FROM ELECTRON STORAGE-RINGS [J].
DUNCAN, WD ;
WILLIAMS, GP .
APPLIED OPTICS, 1983, 22 (18) :2914-2923
[5]   Synchrotron infrared microspectrometry applied to petrography in micrometer-scale range: Fluid chemical analysis and mapping [J].
Guilhaumou, N ;
Dumas, P ;
Carr, GL ;
Williams, GP .
APPLIED SPECTROSCOPY, 1998, 52 (08) :1029-1034
[6]   Highly resolved chemical imaging of living cells by using synchrotron infrared microspectrometry [J].
Jamin, N ;
Dumas, P ;
Moncuit, J ;
Fridman, WH ;
Teillaud, JL ;
Carr, GL ;
Williams, GP .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1998, 95 (09) :4837-4840
[7]   FOURIER-TRANSFORM INFRARED MICROSCOPIC ANALYSIS WITH SYNCHROTRON-RADIATION - THE MICROSCOPE OPTICS AND SYSTEM PERFORMANCE [J].
REFFNER, JA ;
MARTOGLIO, PA ;
WILLIAMS, GP .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02) :1298-1302
[8]  
1998, CELLULAR MICROBIOLOG, V44