Development of a novel optimized breathhold technique for myocardial T2 measurement in thalassemia

被引:61
作者
He, Taigang
Gatehouse, Peter D.
Anderson, Lisa J.
Tanner, Mark
Keegan, Jennifer
Pennell, Dudley J.
Firmin, David N.
机构
[1] Royal Brompton Hosp, Cardiovasc Magnet Resonance Unit, London SW3 6NP, England
[2] Univ London Imperial Coll Sci Technol & Med, London, England
关键词
cardiovascular magnetic resonance (CMR); myocardial T2; fast spin-echo; black-blood imaging; thalassemia; iron loading;
D O I
10.1002/jmri.20681
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To develop a reproducible fast spin-echo (FSE) technique for accurate myocardial T2 measurement with application to iron overload assessment in thalassemia. Materials and Methods: An FSE sequence was developed to permit acquisition of multiple TE images in one breathhold (BH-FSE). A dynamic black-blood scheme was introduced to better cancel blood signal. A nonselective refocusing train was also adopted to suppress stimulated echoes. The optimized technique was tested on phantoms and then applied to 10 normal volunteers and 10 thalassemia patients. Interstudy reproducibility was measured on all the 20 subjects. Results: The mean difference in T2 values was 1.7% from phantom experiments between BH-FSE and the conventional spin-echo (SE) technique. High contrast BH-FSE images were acquired from human subjects, with minimal stimulated echoes and effective blood suppression (P = 0.0005). The coefficient of variation for interstudy reproducibility was 4.3%. T2 values from thalassemia patients were substantially lower than those from the normal subjects (45.2 +/- 26.1 msec vs. 56.9 +/- 8.4 ms, P = 0.02). Conclusion: The dynamic black-blood T2 sequence is a fast reproducible acquisition that compares favourably with conventional techniques, is robust to motion artifacts, and yields high blood-myocardium contrast. This technique may provide a useful tool in thalassemia and other scenarios requiring myocardial T2 quantification.
引用
收藏
页码:580 / 585
页数:6
相关论文
共 24 条
[1]  
ALDOURI MA, 1990, ACTA HAEMATOL-BASEL, V84, P113
[2]   Comparison of effects of oral deferiprone and subcutaneous desferrioxamine on myocardial iron concentrations and ventricular function in beta-thalassaemia [J].
Anderson, LJ ;
Wonke, B ;
Prescott, E ;
Holden, S ;
Walker, JM ;
Pennell, DJ .
LANCET, 2002, 360 (9332) :516-520
[3]   Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload [J].
Anderson, LJ ;
Holden, S ;
Davis, B ;
Prescott, E ;
Charrier, CC ;
Bunce, NH ;
Firmin, DN ;
Wonke, B ;
Porter, J ;
Walker, JM ;
Pennell, DJ .
EUROPEAN HEART JOURNAL, 2001, 22 (23) :2171-2179
[4]   Noninvasive measurement of iron: report of an NIDDK workshop [J].
Brittenham, GM ;
Badman, DG .
BLOOD, 2003, 101 (01) :15-19
[5]   Long-term outcome of continuous 24-hour deferoxamine infusion via indwelling intravenous catheters in high-risk β-thalassemia [J].
Davis, BA ;
Porter, JB .
BLOOD, 2000, 95 (04) :1229-1236
[6]   FAST SHORT-TAU INVERSION-RECOVERY MR IMAGING [J].
FLECKENSTEIN, JL ;
ARCHER, BT ;
BARKER, BA ;
VAUGHAN, JT ;
PARKEY, RW ;
PESHOCK, RM .
RADIOLOGY, 1991, 179 (02) :499-504
[7]   Corrected equations for susceptibility-induced T2-shortening [J].
Gillis, P ;
Roch, A ;
Brooks, RA .
JOURNAL OF MAGNETIC RESONANCE, 1999, 137 (02) :402-407
[8]   HEPATIC IRON OVERLOAD - QUANTITATIVE MR IMAGING [J].
GOMORI, JM ;
HOREV, G ;
TAMARY, H ;
ZANDBACK, J ;
KORNREICH, L ;
ZAIZOV, R ;
FREUD, E ;
KRIEF, O ;
BENMEIER, J ;
ROTEM, H ;
KUSPET, M ;
ROSEN, P ;
RACHMILEWITZ, EA ;
LOEWENTHAL, E ;
GORODETSKY, R .
RADIOLOGY, 1991, 179 (02) :367-369
[9]   Criteria for analysis of multicomponent tissue T-2 relaxation data [J].
Graham, SJ ;
Stanchev, PL ;
Bronskill, MJ .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (03) :370-378
[10]  
HE T, 2005, J CARDIOV MAGN RESON, V7, P287