Improved image quality and low radiation dose with hybrid iterative reconstruction with 80 kV CT pulmonary angiography

被引:38
作者
Laqmani, Azien [1 ]
Regier, Marc [1 ]
Veldhoen, Simon [1 ]
Backhaus, Alexandra [1 ]
Wassenberg, Felicia [1 ]
Sehner, Susanne [3 ]
Groth, Michael [1 ]
Nagel, Hans-Dieter [2 ]
Adam, Gerhard [1 ]
Henes, Frank O. [1 ]
机构
[1] Univ Med Ctr, Dept Diagnost & Intervent Radiol, D-20246 Hamburg, Germany
[2] Sci & Applicat Oriented Studies & Consulting Radi, D-21244 Buchholz, Germany
[3] Univ Med Ctr, Dept Med Biometry & Epidemiol, D-20246 Hamburg, Germany
关键词
Computed tomography pulmonary angiography; Iterative reconstruction; Radiation dose; Low tube voltage; Pulmonary embolism; LOW-TUBE-VOLTAGE; COMPUTED-TOMOGRAPHY ANGIOGRAPHY; VASCULAR ENHANCEMENT; OBESE-PATIENTS; CHEST CT; EMBOLISM; ALGORITHM; PROTOCOLS; ARTERIES; OPTIMIZATION;
D O I
10.1016/j.ejrad.2014.06.016
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Objectives: To determine the impact of hybrid iterative reconstruction (HIR) on image quality in 80 kV CT pulmonary angiography (CTPA) in comparison to filtered-back-projection (FBP). Methods: Fifty patients (body weight <80 kg) with suspected pulmonary embolism (PE) underwent CTPA at 80 kV (mean CTDIvol, 2.3 mGy; effective dose, 1.2 mSv). The raw data were reconstructed using FBP and three increasing HIR levels. Two radiologists assessed image quality and image noise. Conspicuity of PE was assessed in central, segmental, and subsegmental arteries. CT attenuation of pulmonary arteries, objective image noise (OIN) and contrast-to-noise ratios (CNR) were assessed. Results: With each HIR level, a significant decrease in subjective and objective image noise was achieved with a reduction of OIN up to 46% in comparison with FBP. CNR significantly increased with the application of HIR compared to FBP. Image quality was rated significantly higher at HIR reconstructions in comparison with FBP. Diagnosis of PE was feasible with each data set; however, conspicuity of central and segmental PE significantly improved with the use of HIR. Conclusions: Eighty kilovoltage CTPA with HIR provides improved image quality and conspicuity of pulmonary embolism enabling low dose CTPA protocols close to 1 mSv in patients weighing less than 80 kg. (C) 2014 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:1962 / 1969
页数:8
相关论文
共 31 条
[1]
Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[2]
80-kV Pulmonary CT Angiography With 40 mL of Iodinated Contrast Material in Lean Patients: Comparison of Vascular Enhancement With Iodixanol (320 mg I/mL) and Iomeprol (400 mg I/mL) [J].
Faggioni, Lorenzo ;
Neri, Emanuele ;
Sbragia, Paola ;
Pascale, Rachele ;
D'Errico, Luigia ;
Caramella, Davide ;
Bartolozzi, Carlo .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (06) :1220-1225
[3]
Peripheral pulmonary arteries: How far in the lung does multi-detector row spiral CT allow analysis? [J].
Ghaye, B ;
Szapiro, D ;
Mastora, I ;
Delannoy, V ;
Duhamel, A ;
Remy, J ;
Remy-Jardin, M .
RADIOLOGY, 2001, 219 (03) :629-636
[4]
Impact of Tube Current-Time and Tube Voltage Reduction in 64-Detector-row Computed Tomography Pulmonary Angiography for Pulmonary Embolism in a Porcine Model [J].
Henes, Frank Oliver Gerhard ;
Groth, Michael ;
Begemann, Philipp G. C. ;
Bley, Thorsten ;
Adam, Gerhard ;
Regier, Marc .
JOURNAL OF THORACIC IMAGING, 2012, 27 (02) :113-120
[5]
An approach for the estimation of effective radiation dose at CT in pediatric patients [J].
Huda, W ;
Atherton, JV ;
Ware, DE ;
Cumming, WA .
RADIOLOGY, 1997, 203 (02) :417-422
[6]
Radiation dose to the female breast from 16-MDCT body protocols [J].
Hurwitz, LM ;
Yoshizumi, T ;
Reiman, RE ;
Paulson, EK ;
Frush, DP ;
Nguyen, GT ;
Toncheva, GI ;
Goodman, PC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (06) :1718-1722
[7]
Reduction in radiation and contrast medium dose via optimization of low-kilovoltage CT protocols using a hybrid iterative reconstruction algorithm at 256-slice body CT: Phantom study and clinical correlation [J].
Itatani, R. ;
Oda, S. ;
Utsunomiya, D. ;
Funama, Y. ;
Honda, K. ;
Katahira, K. ;
Morishita, S. ;
Yamamura, S. ;
Namimoto, T. ;
Yamashita, Y. .
CLINICAL RADIOLOGY, 2013, 68 (03) :E128-E135
[8]
Chest CT performed with z-axis modulation: Scanning protocol and radiation dose [J].
Kalra, MK ;
Rizzo, S ;
Maher, MM ;
Halpern, EF ;
Toth, TL ;
Shepard, JAO ;
Aquino, SL .
RADIOLOGY, 2005, 237 (01) :303-308
[9]
Use of a Hybrid Iterative Reconstruction Technique to Reduce Image Noise and Improve Image Quality in Obese Patients Undergoing Computed Tomographic Pulmonary Angiography [J].
Kligerman, Seth ;
Mehta, Dhruv ;
Farnadesh, Mahmmoudreza ;
Jeudy, Jean ;
Olsen, Kathryn ;
White, Charles .
JOURNAL OF THORACIC IMAGING, 2013, 28 (01) :49-59
[10]
Impact of a 4th Generation Iterative Reconstruction Technique on Image Quality in Low-Dose Computed Tomography of the Chest in Immunocompromised Patients [J].
Laqmani, A. ;
Buhk, J. H. ;
Henes, F. O. ;
Klink, T. ;
Sehner, S. ;
von Schultzendorff, H. C. ;
Hammerle, D. ;
Nagel, H. D. ;
Adam, G. ;
Regier, M. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2013, 185 (08) :749-757