Silicone breast implant rupture: Pitfalls of magnetic resonance imaging and relative efficacies of magnetic resonance, mammography, and ultrasound

被引:70
作者
Ikeda, DM
Borofsky, HB
Herfkens, RJ
Sawyer-Glover, AM
Birdwell, RL
Glover, GH
机构
[1] Stanford Univ, Sch Med, Dept Radiol, Stanford, CA 94305 USA
[2] Stanford Univ, Richard Lucas Radiol Sci Lab, Stanford, CA 94305 USA
关键词
D O I
10.1097/00006534-199912000-00016
中图分类号
R61 [外科手术学];
学科分类号
摘要
The objective of this study was to evaluate the relative efficacies of magnetic resonance (MR) imaging, ultrasonography, and mammography in implant rupture detection and to illustrate pitfalls in MR image interpretation. Thirty patients referred by plastic surgeons with suspected breast implant rupture were prospectively evaluated using MR ultrasonography, and mammography. Imaging examinations were interpreted independently and blindly for implant rupture and correlated to operative findings. Surgical correlation in 16 patients (53 percent) with 31 implants showed 13 (42 percent) were intact, 5 (16 percent) had severe gel bleed, and 13 (42 percent) were ruptured. MR sensitivity was 100 percent and specificity was 63 percent. Accuracy for rupture was 81 percent with MR, higher than with ultrasonography and mammography (77 and 59 percent, respectively). We describe a specific pitfall in MR interpretation, the "rat-tail" sign, composed of a medial linear extension of silicone along the chest wall. Seen in eight cases (four intact, three ruptures, one gel bleed), the rat-tail sign may lead to misdiagnosis of implant rupture if seen in isolation. Magnetic resonance imaging is more accurate and sensitive than ultrasonography and mammography in detecting breast implant rupture. We describe a new sign (rat-tail sign) composed of medial compression of the implant simulating silicone extrusion as a potential false-positive MR finding for rupture. This article presents clinical experience with magnetic resonance, mammography, and ultrasound in the diagnosis of implant rupture and defines and illustrates potential pitfalls of MR interpretation, including the new rat-tail sign.
引用
收藏
页码:2054 / 2062
页数:9
相关论文
共 25 条
[1]   MR-IMAGING OF THE BREAST IN PATIENTS WITH SILICONE BREAST IMPLANTS - NORMAL POSTOPERATIVE VARIANTS AND DIAGNOSTIC PITFALLS [J].
BERG, WA ;
ANDERSON, ND ;
ZERHOUNI, EA ;
CHANG, BW ;
KUHLMAN, JE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 163 (03) :575-578
[2]   SINGLE-LUMEN AND DOUBLE-LUMEN SILICONE BREAST IMPLANT INTEGRITY - PROSPECTIVE EVALUATION OF MR AND US CRITERIA [J].
BERG, WA ;
CASKEY, CI ;
HAMPER, UM ;
KUHLMAN, JE ;
ANDERSON, ND ;
CHANG, BW ;
SHETH, S ;
ZERHOUNI, EA .
RADIOLOGY, 1995, 197 (01) :45-52
[3]   BREAST AUGMENTATION - A RISK FACTOR FOR BREAST-CANCER [J].
BERKEL, H ;
BIRDSELL, DC ;
JENKINS, H .
NEW ENGLAND JOURNAL OF MEDICINE, 1992, 326 (25) :1649-1653
[4]  
BERKEL HJ, 1992, CAN MED ASSOC J, V147, P1127
[5]   BREAST IMPLANT RUPTURE - DIAGNOSIS WITH US [J].
CASKEY, CI ;
BERG, WA ;
ANDERSON, ND ;
SHETH, S ;
CHANG, BW ;
HAMPER, UM .
RADIOLOGY, 1994, 190 (03) :819-823
[6]   Diagnosis of silicone gel breast implant rupture by ultrasonography [J].
Chung, KC ;
Wilkins, EG ;
Beil, RJ ;
Helvie, MA ;
Ikeda, DM ;
Oneal, RM ;
Forrest, ME ;
Smith, DJ .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1996, 97 (01) :104-109
[7]   SILICONE BREAST IMPLANTS - US EVALUATION [J].
DEBRUHL, ND ;
GORCZYCA, DP ;
AHN, CY ;
SHAW, WW ;
BASSETT, LW .
RADIOLOGY, 1993, 189 (01) :95-98
[8]   SCREENING MAMMOGRAPHY IN 350 WOMEN WITH BREAST IMPLANTS - PREVALENCE AND FINDINGS OF IMPLANT COMPLICATIONS [J].
DESTOUET, JM ;
MONSEES, BS ;
OSER, RF ;
NEMECEK, JR ;
YOUNG, VL ;
PILGRAM, TK .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1992, 159 (05) :973-978
[9]   DIAGNOSIS OF BREAST IMPLANT RUPTURE - IMAGING FINDINGS AND RELATIVE EFFICACIES OF IMAGING TECHNIQUES [J].
EVERSON, LI ;
PARANTAINEN, H ;
DETLIE, T ;
STILLMAN, AE ;
OLSON, PN ;
LANDIS, G ;
FOSHAGER, MC ;
CUNNINGHAM, B ;
GRIFFITHS, HJ .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 163 (01) :57-60
[10]   AUGMENTATION MAMMAPALSTY - NORMAL AND ABNORMAL FINDINGS WITH MAMMOGRAPHY AND ULTRASOUND [J].
GANOTT, MA ;
HARRIS, KM ;
ILKHANIPOUR, ZS ;
COSTAGRECO, MA .
RADIOGRAPHICS, 1992, 12 (02) :281-295