Stripping of failing haemodialysis catheters using the Ampltaz gooseneck snare

被引:38
作者
Rockall, AG
Harris, A
Wetton, CWN
Taube, D
Gedroyc, W
AlKutoubi, MA
机构
[1] ST MARYS HOSP,DEPT RADIOL,LONDON,ENGLAND
[2] ST MARYS HOSP,DEPT NEPHROL,LONDON,ENGLAND
关键词
D O I
10.1016/S0009-9260(97)80255-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The development of a fibrin sheath at the tip of a long-term haemodialysis catheter mag lead to deteriorating blood flow rates, resulting in inadequate haemodialysis. Restoration of functional patency has been described using the technique of percutaneous fibrin sheath stripping (PFSS) using a wire snare de rice. Our purpose was to assess this technique within an established renal vascular access ser, ice. All catheters referred for the assessment of low blood flow rates on dialysis were considered for the procedure. Initial fluoroscopic assessment of the catheter was followed by stripping of the tip of the catheter using a gooseneck snare, Seventeen of 22 well-positioned catheters undergoing a single PFSS attempt were restored to function with a median prolongation of patency of 1.25 months, Two catheters underwent a second PFSS procedure providing additional patency. Nineteen of a total of 24 (79%) PFSS procedures successfully restored catheter function, Seven catheters with poorly positioned tips or a kink were not restored to functional patency using PFSS, Two patients developed a puncture site haematoma, No patient developed symptoms of pulmonary emoblism, In conclusion, PFSS restored function in 79% of attempts in well positioned catheters, The method is technically straightforward, with a low complication rate and has become a routine part of the renal vascular access service.
引用
收藏
页码:616 / 620
页数:5
相关论文
共 26 条
[21]  
Shaffer D, 1992, ASAIO J, V38, P55, DOI 10.1097/00002480-199201000-00013
[22]  
SWARTZ RD, 1994, J AM SOC NEPHROL, V4, P1719
[23]   MAINTAINING THE PATENCY OF DOUBLE-LUMEN SILASTIC JUGULAR CATHETERS FOR HEMODIALYSIS [J].
ULDALL, R ;
BESLEY, ME ;
THOMAS, A ;
SALTER, S ;
NUEZCA, LA ;
VAS, M .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1993, 16 (01) :37-40
[24]   VASCULAR ACCESS FOR HEMODIALYSIS [J].
VANHOLDER, R ;
RINGOIR, S .
ARTIFICIAL ORGANS, 1994, 18 (04) :263-265
[25]   SUCCESSFUL SYSTEMIC FIBRINOLYSIS IN PERMANENT HEMODIALYSIS CATHETER OBSTRUCTION [J].
VOGT, K ;
TILLMANN, U ;
BLUMBERG, A .
NEPHRON, 1987, 45 (02) :174-175
[26]   FATAL HYDROTHORAX ASSOCIATED WITH SUBCLAVIAN VEIN CATHETERIZATION FOR HEMODIALYSIS [J].
ZEIEN, LB ;
NOGUCHI, TT .
AMERICAN JOURNAL OF FORENSIC MEDICINE AND PATHOLOGY, 1992, 13 (04) :326-328