Comparison of the performance of different ICP-MS instruments on the measurement of Cu in a water sample by ICP-IDMS

被引:22
作者
Diemer, J [1 ]
Quétel, CR [1 ]
Taylor, PDP [1 ]
机构
[1] European Commiss Joint Res Ctr, Inst Reference Mat & Measurements, B-2440 Geel, Belgium
关键词
D O I
10.1039/b201443b
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Three different types of ICP-MS instrument, a quadrupole, a single collector magnetic sector and a multicollector magnetic sector instrument, were compared for the measurement of Cu in a water sample by ICP-IDMS. Identical samples were run on the different instruments and various dilution levels were tested. Criteria for comparison were the total combined uncertainties of the isotope ratio measurements and of the amount content measurements. The influence of the different modes of detection (e.g., pulse counting/analog mode detection and single collection/multiple collection) was evaluated. Sources of uncertainty related to the instrument included the repeatability of the measurements, the repeatability of the correction for mass discrimination, the correction for detector dead time effects, the correction for instrumental background and the correction for isobaric interference. External sources of uncertainty, e. g. the uncertainty of the certified reference material (CRM) for correction of mass discrimination, were evaluated as well. Whereas in many cases sample preparation represents a major source of uncertainty, this did not apply here since the sample did not require particular treatment. Total combined uncertainties were calculated according to the ISO/GUM and EURACHEM guidelines. Excellent agreement was observed between all the measurement results obtained with the different instruments and instrumental settings under investigation. Relative combined uncertainties between 0.4 and 0.07% (k=1) were obtained for "absolute'' isotope ratio measurements (isotope ratios directly SI-traceable via a CRM) on a copper blend used for IDMS. For the copper amount content determined by ICP-IDMS, overall uncertainties between 1.0% and 0.45% (k=2) could be achieved, depending on the type of instrument and the measurement conditions used.
引用
收藏
页码:1137 / 1142
页数:6
相关论文
共 23 条
  • [1] [Anonymous], 1995, GUID EXPR UNC MEAS
  • [2] [Anonymous], 1988, INORGANIC MASS SPECT
  • [3] [Anonymous], ISOTOPE DILUTION MAS
  • [4] Sources of uncertainty in isotope ratio measurements by inductively coupled plasma mass spectrometry
    Appelblad, PK
    Rodushkin, I
    Baxter, DC
    [J]. ANALYTICAL CHEMISTRY, 2001, 73 (13) : 2911 - 2919
  • [5] De Bievre P., 1993, ANAL PROC, V30, P328, DOI [10.1039/AP9933000328, DOI 10.1039/AP9933000328]
  • [6] ISOTOPE-DILUTION MASS-SPECTROMETRY - WHAT CAN IT CONTRIBUTE TO ACCURACY IN TRACE ANALYSIS
    DEBIEVRE, P
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 337 (07): : 766 - 771
  • [7] SI-traceable certification of the amount content of cadmium below the ng g-1 level in blood samples by isotope dilution ICP-MS applied as a primary method of measurement
    Diemer, J
    Vogl, J
    Quétel, CR
    Linsinger, T
    Taylor, PDP
    Lamberty, A
    Pauwels, J
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 370 (05) : 492 - 498
  • [8] DIEMER J, IN PRESS ANAL BIOANA
  • [9] *EURACHEM, 1995, QUANT UNC AN MEAS
  • [10] Determining picogram quantities of uranium in urine by isotope dilution inductively coupled plasma mass spectrometry.: Comparison with α-spectrometry
    Haldimann, M
    Baduraux, M
    Eastgate, A
    Froidevaux, P
    O'Donovan, S
    Von Gunten, D
    Zoller, O
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (12) : 1364 - 1369