Novel variable volume injector for performing sample introduction in a miniaturised isotachophoresis device

被引:23
作者
Baldock, SJ
Fielden, PR
Goddard, NJ
Kretschmer, HR
Prest, JE
Brown, BJT
机构
[1] Univ Manchester, Dept Instrumentat & Analyt Sci, Manchester M60 1QD, Lancs, England
[2] Siemens AG, Zent Abt Tech, D-13623 Berlin, Germany
基金
英国工程与自然科学研究理事会;
关键词
isotachophoresis; miniaturization; injection methods; instrumentation; chip technology;
D O I
10.1016/j.chroma.2004.05.062
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A microdevice design furnished with a novel sample injector, capable of delivering variable volume samples, for miniaturised isotachophoretic separations is presented. Micromachining by direct milling was used to realise two flow channel network designs on poly(methyl methacrylate) chips. Both designs comprised a wide bore sample channel interfaced, via a short connection channel, to a narrow bore separation channel. Superior injection performance was observed with a connection channel angled at 45degrees to the separation channel compared to a device using a channel angled at 90degrees. Automated delivery of electrolytes to the microdevice was demonstrated with both hydrostatic pumping and syringe pumps; both gave reproducible sample injection. A range of different sampling strategies were investigated. Isotachophoretic separations of model analytes (metal ions and an anionic dye) demonstrated the potential of the device. Separations of ten metal cations were achieved in under 475 s. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:181 / 188
页数:8
相关论文
共 70 条
[1]  
Alarie JP, 2001, ELECTROPHORESIS, V22, P312, DOI 10.1002/1522-2683(200101)22:2<312::AID-ELPS312>3.0.CO
[2]  
2-3
[3]  
Alarie JP, 2000, ELECTROPHORESIS, V21, P100, DOI 10.1002/(SICI)1522-2683(20000101)21:1<100::AID-ELPS100>3.3.CO
[4]  
2-G
[5]  
Attiya S, 2001, ELECTROPHORESIS, V22, P318, DOI 10.1002/1522-2683(200101)22:2<318::AID-ELPS318>3.0.CO
[6]  
2-G
[7]   A chip-based electrophoresis system with electrochemical detection and hydrodynamic injection [J].
Backofen, U ;
Matysik, FM ;
Lunte, CE .
ANALYTICAL CHEMISTRY, 2002, 74 (16) :4054-4059
[8]   Pressure pinched injection of nanolitre volumes in planar micro-analytical devices [J].
Bai, XX ;
Lee, HJ ;
Rossier, JS ;
Reymond, F ;
Schafer, H ;
Wossner, M ;
Girault, HH .
LAB ON A CHIP, 2002, 2 (01) :45-49
[9]  
Baldock S. J., 2000, Proceedings. MICRO.tec 2000. VDE World Microtechnologies Congress, P615
[10]   Integrated moulded polymer electrodes for performing conductivity detection on isotachophoresis microdevices [J].
Baldock, SJ ;
Fielden, PR ;
Goddard, NJ ;
Prest, JE ;
Brown, BJT .
JOURNAL OF CHROMATOGRAPHY A, 2003, 990 (1-2) :11-22