Two hundred and five patients referred for evaluation of platelet functions and 126 healthy controls were tested with the PFA-100 instrument. A cut-off value of 150 s for collagen/epinephrine (CEPI) closure time (CT) produced most acceptable sensitivity (90%), specificity (85.2%), and positive (82.6%) and negative (91.6%) predictivity values for screening of platelet function disorders and von Willebrand disease (vWD). All patients with vWD and Glanzmann thrombasthenia could be detected by PFA-100. Both CEPI and collagen/adenosine diphosphate (CADP) CTs were elevated in all of these cases. Sensitivity of the device was 81.6% for patients with platelet secretion defects. CADP CT was normal in 63.9% of the patients in this subgroup. Specificity (47%) and positive predictivity (57%) of the instrument were diminished in patients with low hemoglobin concentrations. Depending on the results, an algorithm was developed for screening of platelet function disorders and vWD with PFA-100.
引用
收藏
页码:349 / 353
页数:5
相关论文
共 14 条
[1]   Evaluation of platelet function with the PFA-100 system in patients with congenital defects of platelet secretion [J].
Cattaneo, M ;
Lecchi, A ;
Agati, B ;
Lombardi, R ;
Zighetti, ML .
THROMBOSIS RESEARCH, 1999, 96 (03) :213-217
[2]  
Cattaneo M, 1999, THROMB HAEMOSTASIS, V82, P35
[3]  
Escolar G, 1999, HAEMATOLOGICA, V84, P614
[4]  
Favaloro EJ, 1999, AM J HEMATOL, V62, P165, DOI 10.1002/(SICI)1096-8652(199911)62:3<165::AID-AJH6>3.0.CO
[5]  
2-C
[6]   Utility of the PFA-100® for assessing bleeding disorders and monitoring therapy:: a review of analytical variables, benefits and limitations [J].
Favaloro, EJ .
HAEMOPHILIA, 2001, 7 (02) :170-179
[7]   Screening for von Willebrand disease with a new analyzer using high shear stress: A study of 60 cases [J].
Fressinaud, E ;
Veyradier, A ;
Truchaud, F ;
Martin, I ;
Boyer-Neumann, C ;
Trossaert, M ;
Meyer, D .
BLOOD, 1998, 91 (04) :1325-1331
[8]   Comparison of PFA-100 closure time and template bleeding time of patients with inherited disorders causing defective platelet function [J].
Kerényi, A ;
Schlammadinger, A ;
Ajzner, É ;
Szegedi, I ;
Kiss, C ;
Pap, Z ;
Boda, Z ;
Muszbek, L .
THROMBOSIS RESEARCH, 1999, 96 (06) :487-492
[9]  
LIND SE, 1991, BLOOD, V77, P2547
[10]  
Mammen EF, 1998, SEMIN THROMB HEMOST, V24, P195, DOI 10.1055/s-2007-995840