Concurrent validity and reliability of a novel wireless inertial measurement system to assess trunk movement

被引:77
作者
Bauer, C. M. [1 ,2 ]
Rast, F. M. [1 ]
Ernst, M. J. [1 ]
Kool, J. [1 ]
Oetiker, S. [1 ]
Rissanen, S. M. [3 ]
Suni, J. H. [4 ]
Kankaanpaa, M. [2 ,5 ]
机构
[1] Zurich Univ Appl Sci, Inst Physiotherapy, Dept Hlth, CH-8400 Winterthur, Switzerland
[2] Univ Tampere, Sch Med, FI-33014 Tampere, Finland
[3] Univ Eastern Finland, Dept Appl Phys, Kuopio 70211, Finland
[4] UKK Inst Hlth Promot Res, Tampere 33500, Finland
[5] Tampere Univ Hosp, Dept Phys & Rehabil Med, Tampere 33521, Finland
基金
芬兰科学院;
关键词
Generalizability-theory; Movement disorders; Reproducibility of results; Biomechanical phenomena; LUMBAR SPINE RANGE; BACK-PAIN; POSTURAL CONTROL; PRESSURE; MOTION; INDIVIDUALS; PERFORMANCE; SENSORS;
D O I
10.1016/j.jelekin.2015.06.001
中图分类号
Q189 [神经科学];
学科分类号
071006 [神经生物学];
摘要
Introduction: Assessment of movement dysfunctions commonly comprises trunk range of motion (ROM), movement or control impairment (MCI), repetitive movements (RM), and reposition error (RE). Inertial measurement unit (IMU)-systems could be used to quantify these movement dysfunctions in clinical settings. The aim of this study was to evaluate a novel IMU-system when assessing movement dysfunctions in terms of concurrent validity and reliability. Methods: The concurrent validity of the IMU-system was tested against an optoelectronic system with 22 participants. The reliability of 14 movement dysfunction tests were analysed using generalizability theory and coefficient of variation, measuring 24 participants in seven trials on two days. Results: The IMU-system provided valid estimates of trunk movement in the primary movement direction when compared to the optoelectronic system. Reliability varied across tests and variables. On average, ROM and RM were more reliable, compared to MCI and RE tests. Discussion: When compared to the optoelectronic system, the IMU-system is valid for estimates of trunk movement in the primary movement direction. Four ROM, two MCI, one RM, and one RE test were identified as reliable and should be studied further for inter-subject comparisons and monitoring changes after an intervention. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:782 / 790
页数:9
相关论文
共 40 条
[1]
Reliability of a Measure of Total Lumbar Spine Range of Motion in Individuals with Low Back Pain [J].
Al Zoubi, Fadi M. ;
Preuss, Richard A. .
JOURNAL OF APPLIED BIOMECHANICS, 2013, 29 (06) :670-677
[2]
[Anonymous], 2008, HLTH MEASUREMENT SCA, DOI DOI 10.1093/ACPROF:OSO/9780199231881.001.0001
[3]
[Anonymous], 2001, Generalizability Theory
[4]
Carter R, 2005, REHABILITATION RES P
[5]
Falls in Parkinson's Disease: Kinematic Evidence for Impaired Head and Trunk Control [J].
Cole, Michael H. ;
Silburn, Peter A. ;
Wood, Joanne M. ;
Worringham, Charles J. ;
Kerr, Graham K. .
MOVEMENT DISORDERS, 2010, 25 (14) :2369-2378
[6]
Intrasession reliability of the "center of pressure minus center of mass" variable of postural control in the healthy elderly [J].
Corriveau, H ;
Hébert, R ;
Prince, F ;
Raîche, M .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2000, 81 (01) :45-48
[7]
A new technique for determining 3-D joint angles: the tilt/twist method [J].
Crawford, NR ;
Yamaguchi, GT ;
Dickman, CA .
CLINICAL BIOMECHANICS, 1999, 14 (03) :153-165
[8]
The use of inertial sensors system for human motion analysis [J].
Cuesta-Vargas, Antonio I. ;
Galan-Mercant, Alejandro ;
Williams, Jonathan M. .
PHYSICAL THERAPY REVIEWS, 2010, 15 (06) :462-473
[9]
When to use agreement versus reliability measures [J].
de Vet, Henrica C. W. ;
Terwee, Caroline B. ;
Knol, Dirk L. ;
Bouter, Lex M. .
JOURNAL OF CLINICAL EPIDEMIOLOGY, 2006, 59 (10) :1033-1039
[10]
Deterministic accessory spinal movement in functional tasks characterizes individuals with low back pain [J].
Dideriksen, J. L. ;
Gizzi, L. ;
Petzke, F. ;
Falla, D. .
CLINICAL NEUROPHYSIOLOGY, 2014, 125 (08) :1663-1668