Electron contamination from the lead cutout used in kilovoltage radiotherapy

被引:20
作者
Lee, CHM [1 ]
Chan, KKD [1 ]
机构
[1] Pamela Youde Nethersoe Eastern Hosp, Dept Med Phys, Chaiwan, Peoples R China
关键词
D O I
10.1088/0031-9155/45/1/301
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In kilovoltage radiotherapy, low-energy electrons are generated on the lead cutout plate frequently used to shape the x-ray field. They can cause surface dose enhancement and affect the percentage depth dose (PDD) and cutout factor when these parameters are measured with a parallel-plate chamber. Although many previous studies concerning a similar surface dose enhancement have suggested using a thin electron filter to eliminate the effect of these low-energy electrons, no details have been given regarding the adequacy of the filter. This paper describes the procedures required to test an electron filter for our Darpec 2000 superficial x-ray machine. A piece of plain paper was found to be an adequate electron Biter, which could block all the low-energy electrons without affecting the PDD and cutout factor measurement. The surface dose enhancement caused by the presence of the cutout plate was studied with such a filter. The enhancement was found to increase with x-ray energy and decrease with cutout size as well as applicator size. At the highest x-ray energy in this study (120 kV, HVL 8.6 mmAl), the enhancement ranges from less than 1% for a 8 cm diameter cutout to a maximum of 14% for a 2 cm diameter cutout with a 10 cm applicator. For a 5 cm diameter applicator it ranges from about 1% to about 11%. The difference between the measured and predicted cutout factor displays similar characteristics. Since the low-energy electrons do not generate any useful treatment dose but their presence can lead to potential dosimetry problems, we suggest that a filter tested with the procedure described in this paper should be used when percentage depth dose and cutout factor for kilovoltage x-ray are measured with a parallel-plate chamber.
引用
收藏
页码:1 / 8
页数:8
相关论文
共 10 条
[1]  
[Anonymous], 1996, Phys Med Biol, V41, P2605
[2]  
ATTIX F. H., 1986, Introduction to radiological physics and radiation dosimetry
[3]   Forward dose perturbation at high atomic number interfaces in kilovoltage x-ray beams [J].
Das, IJ .
MEDICAL PHYSICS, 1997, 24 (11) :1781-1787
[4]   Broad beam attenuation of kilovoltage photon beams: effect of ion chambers [J].
Das, IJ .
BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (841) :68-73
[5]   COMPLICATIONS IN LOW-ENERGY X-RAY DOSIMETRY CAUSED BY ELECTRON CONTAMINATION [J].
KLEVENHAGEN, SC ;
DSOUZA, D ;
BONNEFOUX, I .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (08) :1111-1116
[6]   Measurement of percentage depth dose and lateral beam profile for kilovoltage x-ray therapy beams [J].
Li, XA ;
Ma, CM ;
Salhani, D .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (12) :2561-2568
[7]   SURFACE DOSE IN INTRACAVITARY ORTHOVOLTAGE RADIOTHERAPY [J].
PODGORSAK, MB ;
SCHREINER, LJ ;
PODGORSAK, EB .
MEDICAL PHYSICS, 1990, 17 (04) :635-640
[8]   CONSIDERATIONS FOR SUPERFICIAL PHOTON DOSIMETRY [J].
STERN, RL ;
KUBO, HD .
MEDICAL PHYSICS, 1995, 22 (09) :1469-1470
[9]  
WHITTON JT, 1973, HEALTH PHYS, V24, P1, DOI 10.1097/00004032-197301000-00001
[10]  
1996, CTR AXIX DEPT DOSE B, V25