Measurement of percentage depth dose and lateral beam profile for kilovoltage x-ray therapy beams

被引:34
作者
Li, XA
Ma, CM
Salhani, D
机构
[1] Ottawa Reg Canc Ctr, Dept Med Phys, Ottawa, ON K1Y 4K7, Canada
[2] Stanford Univ, Dept Radiat Oncol, Stanford, CA 94305 USA
关键词
D O I
10.1088/0031-9155/42/12/019
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this work, nine commonly used dosimetry detectors have been investigated to determine suitable relative dosimeters for kilovoltage x-ray beams. By comparison with the Monte Carlo calculated data, it was determined that for the detectors studied the PTW N23342, Markus and NACP parallel-plate chambers are more suitable for the measurement of percentage depth dose (PDD) data for this beam quality range with an uncertainty of about 3%. A diode detector may be used to measure the PDD for the 100 kVp beam, but it is not suitable for higher energies (300 kVp). The Capintec parallel-plate chamber may be adequate for medium-energy photons, but it has a slightly higher uncertainty for low-energy x-rays (100 kVp). For the measurement of beam profiles, diode and film yield incorrect profile tails, which can be corrected using the RK ionization chamber.
引用
收藏
页码:2561 / 2568
页数:8
相关论文
共 23 条
[1]  
Aldrich J E, 1992, Med Dosim, V17, P69
[2]  
[Anonymous], P IAEA INT S MEAS AS
[3]   Performance characteristics of the Pantak DXT-300 kilovoltage X-ray treatment machine [J].
Aukett, RJ ;
Thomas, DW ;
Seaby, AW ;
Gittins, JT .
BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (824) :726-734
[4]  
*BJR, 1983, BR J RADIOL S, V17
[5]   THE DEPENDENCE OF MASS-ENERGY ABSORPTION-COEFFICIENT RATIOS ON BEAM SIZE AND DEPTH IN A PHANTOM [J].
CUNNINGHAM, JR ;
WOO, M ;
ROGERS, DWO ;
BIELAJEW, AF .
MEDICAL PHYSICS, 1986, 13 (04) :496-502
[6]   BEAM CHARACTERISTICS OF THE THERAPAX DXT300 ORTHOVOLTAGE THERAPY UNIT [J].
GERIG, L ;
SOUBRA, M ;
SALHANI, D .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (09) :1377-1392
[7]   CENTRAL-AXIS DEPTH-DOSE DATA FOR DIAGNOSTIC-RADIOLOGY [J].
HARRISON, RM .
PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (04) :657-670
[8]  
*IAEA, 1987, 277 IAEA
[9]  
ICRU, 1973, 23 ICRU
[10]  
IPSM, 1991, PHYS MED BIOL, V36, P1027