Characterization of UV-induced radiation damage to Si-based photodiodes

被引:4
作者
Gupta, R [1 ]
Lykke, KR [1 ]
Shaw, PS [1 ]
Dehmer, JL [1 ]
机构
[1] Natl Inst Stand & Technol, Opt Technol Div, Gaithersburg, MD 20899 USA
来源
ULTRAVIOLET ATMOSPHERIC AND SPACE REMOTE SENSING: METHODS AND INSTRUMENTATION II | 1999年 / 3818卷
关键词
photodiodes; silicon; ultraviolet; cryogenic radiometer; synchrotron radiation;
D O I
10.1117/12.364158
中图分类号
V [航空、航天];
学科分类号
08 ; 0825 ;
摘要
We have made direct measurements of the internal quantum efficiency and the reflectivity of W-damaged silicon photodiodes in the spectral range of 125 nm to 320 nm. The above quantities, coupled with absolute spectral responsivities, may yield unique information leading to the identification of the mechanisms responsible for the degradation of performance of the silicon photodiodes in the ultraviolet. The measurements were made using synchrotron radiation from the NIST synchrotron ultraviolet radiation facility (SURF) and an absolute cryogenic radiometer as a primary standard detector.
引用
收藏
页码:27 / 33
页数:7
相关论文
共 16 条
[1]  
BOOKER RL, 1977, APPL OPTICS, V16, P1539
[2]  
BUDDE W, 1983, OPTICAL RAD MEASUREM, V4, P244
[3]   Absolute silicon photodiodes for 160 nm to 254 nm photons [J].
Canfield, LR ;
Vest, RE ;
Korde, R ;
Schmidtke, H ;
Desor, R .
METROLOGIA, 1998, 35 (04) :329-334
[4]  
CANFIELD LR, 1989, APPL OPT, V28
[5]   A PHYSICAL BASIS FOR THE EXTRAPOLATION OF SILICON PHOTODIODE QUANTUM EFFICIENCY INTO THE ULTRAVIOLET [J].
DURANT, NM ;
FOX, NP .
METROLOGIA, 1993, 30 (04) :345-350
[6]   TRAP DETECTORS AND THEIR PROPERTIES [J].
FOX, NP .
METROLOGIA, 1991, 28 (03) :197-202
[7]   ONE GIGARAD PASSIVATING NITRIDED OXIDES FOR 100-PERCENT INTERNAL QUANTUM EFFICIENCY SILICON PHOTODIODES [J].
KORDE, R ;
CABLE, JS ;
CANFIELD, LR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (06) :1655-1659
[8]  
KORDE R, 1989, APPL OPT, V26
[9]   Characterization of photodiodes as transfer detector standards in the 120 nm to 600 nm spectral range [J].
Kuschnerus, P ;
Rabus, H ;
Richter, M ;
Scholze, F ;
Werner, L ;
Ulm, G .
METROLOGIA, 1998, 35 (04) :355-362
[10]   First results with the new PTB cryogenic radiometer for the vacuum-ultraviolet spectral range [J].
LauFrambs, A ;
Kroth, U ;
Rabus, H ;
Tegeler, E ;
Ulm, G ;
Wende, B .
METROLOGIA, 1996, 32 (06) :571-574