First results with the new PTB cryogenic radiometer for the vacuum-ultraviolet spectral range

被引:18
作者
LauFrambs, A
Kroth, U
Rabus, H
Tegeler, E
Ulm, G
Wende, B
机构
关键词
D O I
10.1088/0026-1394/32/6/34
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In the radiometry laboratory of the Physikalisch-Technische Bundesanstalt (PTB) at the Berlin electron storage ring BESSY, a new cryogenic electrical-substitution radiometer with a cavity absorber is now in operation as the primary detector standard for the vacuum-ultraviolet spectral range. The instrument is optimized for use with narrowband synchrotron radiation, as dispersed by a monochromator. It is capable of measuring spectral radiant powers of a few microwatts with a resolution of 1,5 nW and a relative uncertainty of 0,2%. The cryogenic radiometer has been employed at a new detector-calibration facility for the normal-incidence spectral range. Calibrations of Hamamatsu S1337 and G1127 photodiodes have been performed in the 170 nm to 500 nm spectral range with a relative uncertainty of the spectral responsivity less than 1%. At four laser wavelengths, agreement was found with a calibration against the laser-operated cryogenic radiometer of the PTB at Berlin.
引用
收藏
页码:571 / 574
页数:4