Accuracy of atrial tachyarrhythmia detection in implantable devices with arrhythmia therapies

被引:137
作者
Purerfellner, H
Gillis, AM
Holbrook, R
Hettrick, DA
机构
[1] Medtronic Inc, Minneapolis, MN 55432 USA
[2] Alberta Heritage Fdn Med Res, Calgary, AB, Canada
[3] Foothills Med Ctr, Calgary, AB, Canada
[4] Univ Calgary, Calgary, AB T2N 1N4, Canada
[5] St Elisabeths Sisters Hosp, Linz, Austria
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2004年 / 27卷 / 07期
关键词
atrial fibrillation; pacemakers; implantable defibrillators; diagnostics;
D O I
10.1111/j.1540-8159.2004.00569.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The clinical application of atrial tachyarrhythmia (AT) episode data stored by implantable devices is presently limited by the high proportion of inappropriate detection. We quantified the percentage of inappropriate AT detection in two implantable devices with AT diagnostics and therapies via meta-analysis of stored AT episodes from a number of clinical trials. The AT500 and GEM III AT, contain dual chamber logic to discriminate AT from ventricular tachycardia and far-field R wave (FFRW) oversensing using dual chamber bipolar electrograms. A subset of data from four clinical trials of 1,142 patients was considered. Manual analysis was performed on 21,553 stored episodes with atrial EGM and marker channel from 409 patients with stored episodes and the market-released device detection configuration. The percentage of episodes with inappropriate detection and termination was evaluated and compared between septal and nonseptal lead locations. The percentage of inappropriately detected episodes receiving ATP therapy was also determined. The percentage of episodes appropriately detected and the percentage of net episode duration (i.e., burden) recorded by the device were also determined from a separate analysis of 24-hour Holter recordings from a subset of 40 patients from one trial. Adjusted estimates of the percentage of inappropriate detection were 95.3% (93.5-96.7; 95% CI) for AT500 and 95.7% (84.3-98.9) for GEM III AT. Inappropriate detection was primarily due to FFRW oversensing or brief runs of premature atrial contractions (PACs). The device detected 100% of the sustained atrial arrhythmia episodes and 95.3% (range 76.1-99.9) of the net AT duration observed on the Holter recordings. AT detection was not influenced by atrial lead location. Appropriate detection of normal sinus rhythm at episode termination was 83.7% (80.7-86.3) for AT500 and 92.1% (84.5-96.2) for GEM III AT Accurate detection and discrimination of FFRWs validates the reliability of AT diagnostic data and decreases the risk of inappropriate device therapy.
引用
收藏
页码:983 / 992
页数:10
相关论文
共 23 条
  • [1] BAYES T, 1958, BIOMETRIKA, V45, P296
  • [2] Impact and prevention of far-field sensing in fallback mode switches
    Bordacher, P
    Cazeau, S
    Graindorg, L
    Ritter, P
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (01): : 206 - 209
  • [3] A new pacemaker algorithm for the treatment of atrial fibrillation - Results of the Atrial Dynamic Overdrive Pacing Trial (ADOPT)
    Carlson, MD
    Ip, J
    Messenger, J
    Beau, S
    Kalbfleisch, S
    Gervais, P
    Cameron, DA
    Duran, A
    Val-Mejias, J
    Mackall, J
    Gold, M
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 42 (04) : 627 - 633
  • [4] Contributions of high resolution electrograms memorized by DDDR pacemakers in the interpretation of arrhythmic events
    Defaye, P
    Leclercq, JF
    Guilleman, D
    Scanu, P
    Hazard, JR
    Fatemi, M
    Boursier, M
    Lambiez, M
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (01): : 214 - 220
  • [5] Management of atrial tachyarrhythmias:: Benefits of pacemaker diagnostics
    Fauchier, L
    Briand, F
    Soto, FX
    Quennelle, F
    Lévy, J
    Darmon, JP
    Lellouche, D
    Lavergne, T
    Poret, P
    Pelade, C
    Babuty, D
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (01): : 233 - 238
  • [6] Atrial therapies reduce atrial arrhythmia burden in defibrillator patients
    Friedman, PA
    Dijkman, B
    Warman, EN
    Xia, A
    Mehra, R
    Stanton, MS
    Hammill, SC
    [J]. CIRCULATION, 2001, 104 (09) : 1023 - 1028
  • [7] Safety and efficacy of advanced atrial pacing therapies for atrial tachyarrhythmias in patients with a new implantable dual chamber cardioverter-defibrillator
    Gillis, AM
    Unterberg-Buchwald, C
    Schmidinger, H
    Massimo, S
    Wolfe, K
    Kavaney, DJ
    Otterness, MF
    Hohnloser, SH
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 40 (09) : 1653 - 1659
  • [8] Atrial pacing periablation for prevention of paroxysmal atrial fibrillation
    Gillis, AM
    Wyse, DG
    Connolly, SJ
    Dubuc, M
    Philippon, F
    Yee, R
    Lacombe, P
    Rose, MS
    Kerr, CD
    [J]. CIRCULATION, 1999, 99 (19) : 2553 - 2558
  • [9] Incremental programming of atrial anti-tachycardia pacing therapies in bradycardia-indicated patients:: effects on therapy efficacy and atrial tachyarrhythmia burden
    Hügl, B
    Israel, CW
    Unterberg, C
    Lawo, T
    Geller, JC
    Kennis, IM
    Euler, DE
    Koehler, JL
    Hettrick, DA
    [J]. EUROPACE, 2003, 5 (04): : 403 - 409
  • [10] Pacing to prevent atrial fibrillation
    Israel, CW
    Hohnloser, SH
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2003, 14 (09) : S20 - S26