Planar Hall effect sensor for magnetic micro- and nanobead detection

被引:162
作者
Ejsing, L
Hansen, MF
Menon, AK
Ferreira, HA
Graham, DL
Freitas, PP
机构
[1] Tech Univ Denmark, Dept Micro & Nanotechnol, DK-2800 Lyngby, Denmark
[2] Inst Engn Syst & Comp Microsyst & Nanotechnol, P-1000029 Lisbon, Portugal
关键词
D O I
10.1063/1.1759380
中图分类号
O59 [应用物理学];
学科分类号
摘要
Magnetic bead sensors based on the planar Hall effect in thin films of exchange-biased permalloy have been fabricated and characterized. Typical sensitivities are 3 muV/Oe mA. The sensor response to an applied magnetic field has been measured without and with coatings of commercially available 2 mum and 250 nm magnetic beads used for bioapplications (Micromer-M and Nanomag-D, Micromod, Germany). Detection of both types of beads and single bead detection of 2 mum beads is demonstrated, i.e., the technique is feasible for magnetic biosensors. Single 2 mum beads yield 300 nV signals at 10 mA and 15 Oe applied field. (C) 2004 American Institute of Physics.
引用
收藏
页码:4729 / 4731
页数:3
相关论文
共 13 条
  • [1] A biosensor based on magnetoresistance technology
    Baselt, DR
    Lee, GU
    Natesan, M
    Metzger, SW
    Sheehan, PE
    Colton, RJ
    [J]. BIOSENSORS & BIOELECTRONICS, 1998, 13 (7-8) : 731 - 739
  • [2] Detection of a single magnetic microbead using a miniaturized silicon Hall sensor
    Besse, PA
    Boero, G
    Demierre, M
    Pott, V
    Popovic, R
    [J]. APPLIED PHYSICS LETTERS, 2002, 80 (22) : 4199 - 4201
  • [3] The BARC biosensor applied to the detection of biological warfare agents
    Edelstein, RL
    Tamanaha, CR
    Sheehan, PE
    Miller, MM
    Baselt, DR
    Whitman, LJ
    Colton, RJ
    [J]. BIOSENSORS & BIOELECTRONICS, 2000, 14 (10-11) : 805 - 813
  • [4] EJSING L, 2003, P 17 EUR C SOL STAT, P1095
  • [5] Biodetection using magnetically labeled biomolecules and arrays of spin valve sensors (invited)
    Ferreira, HA
    Graham, DL
    Freitas, PP
    Cabral, JMS
    [J]. JOURNAL OF APPLIED PHYSICS, 2003, 93 (10) : 7281 - 7286
  • [6] Ion beam deposition of Mn-Ir spin valves
    Gehanno, V
    Freitas, PP
    Veloso, A
    Ferreira, J
    Almeida, B
    Sousa, JB
    Kling, A
    Soares, JC
    da Silva, MF
    [J]. IEEE TRANSACTIONS ON MAGNETICS, 1999, 35 (05) : 4361 - 4367
  • [7] High sensitivity detection of molecular recognition using magnetically labelled biomolecules and magnetoresistive sensors
    Graham, DL
    Ferreira, HA
    Freitas, PP
    Cabral, JMS
    [J]. BIOSENSORS & BIOELECTRONICS, 2003, 18 (04) : 483 - 488
  • [8] Single magnetic microsphere placement and detection on-chip using current line designs with integrated spin valve sensors: Biotechnological applications
    Graham, DL
    Ferreira, H
    Bernardo, J
    Freitas, PP
    Cabral, JMS
    [J]. JOURNAL OF APPLIED PHYSICS, 2002, 91 (10) : 7786 - 7788
  • [9] Detection of single micron-sized magnetic bead and magnetic nanoparticles using spin valve sensors for biological applications
    Li, GX
    Joshi, V
    White, RL
    Wang, SX
    Kemp, JT
    Webb, C
    Davis, RW
    Sun, SH
    [J]. JOURNAL OF APPLIED PHYSICS, 2003, 93 (10) : 7557 - 7559
  • [10] A DNA array sensor utilizing magnetic microbeads and magnetoelectronic detection
    Miller, MM
    Sheehan, PE
    Edelstein, RL
    Tamanaha, CR
    Zhong, L
    Bounnak, S
    Whitman, LJ
    Colton, RJ
    [J]. JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2001, 225 (1-2) : 138 - 144