Technical Note: Output and energy fluctuations of the tomotherapy Hi-Art helical tomotherapy system

被引:14
作者
Mahan, SL [1 ]
Chase, DJ [1 ]
Ramsey, CR [1 ]
机构
[1] Thompson Canc Survival Ctr, Knoxville, TN 37916 USA
关键词
helical tomotherapy; dose calibration; energy;
D O I
10.1118/1.1763007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The output and energy calibrations for the first clinical Hi-Art 2.0 helical tomotherapy system have been reviewed. Fixed-gantry/fixed-couch and rotational-gantry/fixed-couch measurements were made on a daily basis over a period of 20 weeks to investigate system stability. Static gantry measurements were taken at 10 cm depth in a rectangular stack of Virtual Water at an SSD distance of 90 cm and a field size of 5 x 40 cm. Rotational gantry measurements were taken in a cylindrical phantom Virtual Water phantom for a field size of 5 x 40 cm. The Hi-Art 2.0 system has maintained its calibration to within +/-2% and energy to within +/-1.5% over the initial 20 week period. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:2119 / 2120
页数:2
相关论文
共 7 条
[1]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[2]  
[Anonymous], 1983, MED PHYS, V10, P741
[3]   Clinical helical tomotherapy commissioning dosimetry [J].
Balog, J ;
Olivera, G ;
Kapatoes, J .
MEDICAL PHYSICS, 2003, 30 (12) :3097-3106
[4]   Benchmarking beam alignment for a clinical helical tomotherapy device [J].
Balog, J ;
Mackie, TR ;
Pearson, D ;
Hui, S ;
Paliwal, B ;
Jeraj, R .
MEDICAL PHYSICS, 2003, 30 (06) :1118-1127
[5]  
KHAN FM, 1994, PHYSICS RAD THERAPY, P184
[6]   Tomotherapy [J].
Mackie, TR ;
Balog, J ;
Ruchala, K ;
Shepard, D ;
Aldridge, S ;
Fitchard, E ;
Reckwerdt, P ;
Olivera, G ;
McNutt, T ;
Mehta, M .
SEMINARS IN RADIATION ONCOLOGY, 1999, 9 (01) :108-117
[7]   TOMOTHERAPY - A NEW CONCEPT FOR THE DELIVERY OF DYNAMIC CONFORMAL RADIOTHERAPY [J].
MACKIE, TR ;
HOLMES, T ;
SWERDLOFF, S ;
RECKWERDT, P ;
DEASY, JO ;
YANG, J ;
PALIWAL, B ;
KINSELLA, T .
MEDICAL PHYSICS, 1993, 20 (06) :1709-1719