A quality control (QC) test suitable for routine daily use has been developed for video based electronic portal imaging devices. It provides an objective and quantitative test for acceptable image quality on the basis of the high contrast spatial resolution and the contrast-to-noise ratio (CNR). The test uses a phantom consisting of five sets of high-contrast rectangular bar patterns with spatial frequencies of 0.1, 0.2, 0.25, 0.4, and 0.75 1p/mm. Data obtained during a one month calibration period were used to determine a critical frequency f(c) for the relative square wave modulation transfer function and a critical contrast-to-noise ratio (CNR(c)). Subsequent measurements indicating significant deviations from these critical values result in warning messages to the operator indicating potential problems in system performance. Measurements over a period of two years show that the QC test provides a sensitive indication of imaging performance. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:1237 / 1244
页数:8
相关论文
共 19 条
[1]  
Barnes G. T., 1979, PHYSICS MED IMAGING, P138
[2]   A REVIEW OF ELECTRONIC PORTAL IMAGING DEVICES (EPIDS) [J].
BOYER, AL ;
ANTONUK, L ;
FENSTER, A ;
VANHERK, M ;
MEERTENS, H ;
MUNRO, P ;
REINSTEIN, LE ;
WONG, J .
MEDICAL PHYSICS, 1992, 19 (01) :1-16
[3]   AN OBJECTIVE METHOD FOR EVALUATING ELECTRONIC PORTAL IMAGING DEVICES [J].
DONG, L ;
BOYER, AL .
MEDICAL PHYSICS, 1994, 21 (06) :755-760
[4]   A PRACTICAL METHOD TO MEASURE THE MTF OF CT SCANNERS [J].
DROEGE, RT ;
MORIN, RL .
MEDICAL PHYSICS, 1982, 9 (05) :758-759
[5]   A PRACTICAL METHOD TO ROUTINELY MONITOR RESOLUTION IN DIGITAL IMAGES [J].
DROEGE, RT .
MEDICAL PHYSICS, 1983, 10 (03) :337-343
[6]  
GLUHCHEV G, 1994, PHYS MED BIOL A, V39, P490
[7]   A METHOD AND CONTRAST-DETAIL PHANTOM FOR THE QUANTITATIVE ASSESSMENT OF RADIOTHERAPY PORTAL IMAGING-SYSTEMS [J].
HARRISON, RM ;
LAMBERT, GD .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (796) :384-388
[8]   CLINICAL USE OF ONLINE PORTAL IMAGING FOR DAILY PATIENT TREATMENT VERIFICATION [J].
HERMAN, MG ;
ABRAMS, RA ;
MAYER, RR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (04) :1017-1023
[9]   COMPREHENSIVE QA FOR RADIATION ONCOLOGY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK-GROUP-40 [J].
KUTCHER, GJ ;
COIA, L ;
GILLIN, M ;
HANSON, WF ;
LEIBEL, S ;
MORTON, RJ ;
PALTA, JR ;
PURDY, JA ;
REINSTEIN, LE ;
SVENSSON, GK ;
WELLER, M ;
WINGFIELD, L .
MEDICAL PHYSICS, 1994, 21 (04) :581-618
[10]   DIGITAL CONTRAST ENHANCEMENT FOR ONLINE PORTAL IMAGING [J].
LESZCZYNSKI, KW ;
SHALEV, S .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1989, 27 (05) :507-512