Warning diagnostics for inductively coupled plasma-mass spectrometry

被引:16
作者
Ying, H
Murphy, J
Tromp, JW
Mermet, JM
Salin, ED
机构
[1] McGill Univ, Dept Chem, Montreal, PQ H3A 2K6, Canada
[2] Univ Lyon 1, Sci Analyt Lab, CNRS 5619, F-69622 Villeurbanne, France
基金
加拿大自然科学与工程研究理事会;
关键词
ICP-MS; diagnosis; pattern recognition; elemental analysis;
D O I
10.1016/S0584-8547(00)00144-0
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The performance of an inductively coupled plasma mass spectrometer is substantially impacted by changes in radio frequency power, nebulizer gas flow, and sample liquid flow. A simple diagnostic procedure has been developed to automatically ensure that the quality of results is not compromised by system degradation. The diagnosis procedure was studied in two different stages of instrument operation: the normal operating stage and the warm-up stage. For normal operating stage monitoring, a decision tree was derived with the C4.5 induction engine applied to As-based signals allowing instrument performance to be classified as either normal or abnormal. Abnormal signals were further classified as to their probable cause. Signals present in both deionized water and 0.5% nitric acid spectra are sufficient for diagnosis with an accuracy of 96.0% and 94.0%, respectively, allowing the use of 'blank' solutions to monitor instrument performance. For the warm-up stage diagnosis, a standard solution containing Li, In, Ba and Bi, was used. Stepwise regression was applied as the key factor filter while partial least square (PLS) was adapted to be the classifier. In addition to the modeling test, the importance of the intensities and intensity ratios of different masses was studied. By eliminating the standard element signals from the model, the possibility of using water/nitric acid as test solutions was also studied. The overall prediction accuracy for the warm-up stage is approximately 92% using the standard solution or either the water or nitric acid type blank solutions. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:311 / 326
页数:16
相关论文
共 20 条
[1]  
Beebe KR, 1998, CHEMOMETRICS
[2]   COMPARISON OF PATTERN-RECOGNITION TECHNIQUES FOR SAMPLE CLASSIFICATION USING ELEMENTAL COMPOSITION - APPLICATIONS FOR ICP-AES [J].
BRANAGH, W ;
YU, HN ;
SALIN, ED .
APPLIED SPECTROSCOPY, 1995, 49 (07) :964-970
[3]   System for automatic selection of operating conditions for inductively coupled plasma atomic emission spectroscopy [J].
Branagh, W ;
Whelan, C ;
Salin, ED .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (11) :1307-1315
[4]  
BRANAGH W, 1995, SPECTROSCOPY EUGENE, V10, P20
[5]  
Branagh WA, 1997, CAN J ANAL SCI SPECT, V42, P130
[6]  
BURTON LL, 1993, SPECTROCHIM ACTA B, V48, pE1063
[7]   Nebulizer diagnostics: fundamental parameters, challenges, and techniques on the horizon - Invited lecture [J].
McLean, JA ;
Minnich, MG ;
Iacone, LA ;
Liu, HY ;
Montaser, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (09) :829-842
[8]  
MURPHY J, 1999, EUR WINT C PLASM SPE
[9]   SIMPLE EXPERIMENTS FOR THE CONTROL, THE EVALUATION AND THE DIAGNOSIS OF INDUCTIVELY-COUPLED PLASMA SEQUENTIAL SYSTEMS [J].
POUSSEL, E ;
MERMET, JM ;
SAMUEL, O .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1993, 48 (6-7) :743-755
[10]  
QUINLAN JR, 1993, C4 5 ALGORITHMS MACH