A dual computed tomography linear accelerator unit for stereotactic radiation therapy: A new approach without cranially fixated stereotactic frames

被引:87
作者
Uematsu, M
Fukui, T
Shioda, A
Tokumitsu, H
Takai, K
Kojima, T
Asai, Y
Kusano, S
机构
[1] Department of Radiology, Division of Radiation Oncology, National Defense Medical College, Saitama
[2] Department of Radiology, Division of Radiation Oncology, National Defense Medical College, Tokorozawa, Saitama, 359, 3-2, Namiki
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1996年 / 35卷 / 03期
关键词
radiation therapy; stereotactic; fractionated; treatment unit; linear accelerator; computed tomography;
D O I
10.1016/S0360-3016(96)80022-8
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To perform stereotactic radiation therapy (SRT) without cranially fixated stereotactic frames, we developed a dual computed tomography (CT) linear accelerator (linac) treatment unit. Methods and Materials: This unit is composed of a linac, CT, and motorized table. The linac and CT are set up at opposite ends of the table, which is suitable for both machines. The gantry axis of the linac is coaxial with that of the CT scanner. Thus, the center of the target detected with the CT can be matched easily with the gantry axis of the linac by rotating the table. Positioning is confirmed with the CT for each treatment session. Positioning and treatment errors with this unit were examined by phantom studies. Between August and December 1994, 8 patients with 11 lesions of primary or metastatic brain tumors received SRT with this unit. All lesions were treated with 24 Gy in three fractions to 30 Gy in 10 fractions to the 80% isodose line, with or without conventional external beam radiation therapy. Results: Phantom studies revealed that treatment errors with this unit were within 1 mm after careful positioning. The position was easily maintained using two tiny metallic balls as vertical and horizontal marks. Motion of patients was negligible using a conventional heat-flexible head mold and dental impression. The overall time for a multiple noncoplanar arcs treatment for a single isocenter was less than 1 h on the initial tretment day and usually less than 20 min on subsequent days. Treatment was outpatient-based and well tolerated with no acute toxicities. Satisfactory responses have been documented. Conclusion: Using this treatment unit, multiple fractionated SRT is performed easily and precisely without cranially fixated stereotactic frames.
引用
收藏
页码:587 / 592
页数:6
相关论文
共 13 条
  • [1] Aoki Y, 1987, Radiat Med, V5, P131
  • [2] THE RADIOBIOLOGY OF RADIOSURGERY - RATIONALE FOR DIFFERENT TREATMENT REGIMES FOR AVMS AND MALIGNANCIES
    HALL, EJ
    BRENNER, DJ
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (02): : 381 - 385
  • [3] THE CT SCANNER AS A THERAPY MACHINE
    IWAMOTO, KS
    NORMAN, A
    KAGAN, AR
    WOLLIN, M
    OLCH, A
    BELLOTTI, J
    INGRAM, M
    SKILLEN, RG
    [J]. RADIOTHERAPY AND ONCOLOGY, 1990, 19 (04) : 337 - 343
  • [4] ADAPTATION AND VERIFICATION OF THE RELOCATABLE GILL-THOMAS-COSMAN FRAME IN STEREOTAXIC RADIOTHERAPY
    KOOY, HM
    DUNBAR, SF
    TARBELL, NJ
    MANNARINO, E
    FERARRO, N
    SHUSTERMAN, S
    BELLERIVE, M
    FINN, L
    MCDONOUGH, CV
    LOEFFLER, JS
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 30 (03): : 685 - 691
  • [5] AN ADAPTER FOR COMPUTED TOMOGRAPHY GUIDED STEREOTAXIS
    LAITINEN, LV
    LILIEQUIST, B
    FAGERLUND, M
    ERIKSSON, AT
    [J]. SURGICAL NEUROLOGY, 1985, 23 (06): : 559 - 566
  • [6] CONSENSUS STATEMENT ON STEREOTAXIC RADIOSURGERY QUALITY IMPROVEMENT
    LARSON, DA
    BOVA, F
    EISERT, D
    KLINE, R
    LOEFFLER, J
    LUTZ, W
    MEHTA, M
    PALTA, J
    SCHEWE, K
    SCHULTZ, C
    SHAW, E
    WILSON, F
    LUNSFORD, LD
    ALEXANDER, E
    CHAPMAN, P
    COFFEY, R
    FRIEDMAN, W
    HARSH, G
    MACIUNAS, R
    OLIVIER, A
    STEINBERG, G
    WALSH, J
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (02): : 527 - 530
  • [7] THE RADIOBIOLOGY OF RADIOSURGERY
    LARSON, DA
    FLICKINGER, JC
    LOEFFLER, JS
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (03): : 557 - 561
  • [8] CURRENT RADIOSURGERY PRACTICE - RESULTS OF AN ASTRO SURVEY
    LARSON, DA
    BOVA, F
    EISERT, D
    KLINE, R
    LOEFFLER, J
    LUTZ, W
    MEHTA, M
    PALTA, J
    SCHEWE, K
    SCHULTZ, C
    SHAW, E
    WILSON, JF
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (02): : 523 - 526
  • [9] LOEFFLER JS, 1993, STEREOTACTIC SURG RA, P307
  • [10] TOMOTHERAPY - A NEW CONCEPT FOR THE DELIVERY OF DYNAMIC CONFORMAL RADIOTHERAPY
    MACKIE, TR
    HOLMES, T
    SWERDLOFF, S
    RECKWERDT, P
    DEASY, JO
    YANG, J
    PALIWAL, B
    KINSELLA, T
    [J]. MEDICAL PHYSICS, 1993, 20 (06) : 1709 - 1719