Reproducible product-ion tandem mass spectra on various liquid chromatography/mass spectrometry instruments for the development of spectral libraries

被引:61
作者
Bristow, AWT
Webb, KS
Lubben, AT
Halket, J
机构
[1] LGC Ltd, Teddington TW11 0LY, Middx, England
[2] Bruker Dalton Ltd, Coventry CV4 9GH, W Midlands, England
[3] Univ London Royal Holloway & Bedford New Coll, Egham TW20 0EX, Surrey, England
关键词
D O I
10.1002/rcm.1492
中图分类号
Q5 [生物化学];
学科分类号
071010 [生物化学与分子生物学]; 081704 [应用化学];
摘要
The results of the comparison of product-ion tandem mass (MS/MS) spectra recorded on three ion trap mass spectrometers, a triple quadrupole mass spectrometer and a Fourier transform ion cyclotron resonance mass spectrometer are reported. The spectra were recorded in accordance with a simple experimental protocol, which involved the collision-induced dissociation (CID) attenuation of the abundance of the [M+H](+) ion to between 10 and 50% of its original abundance. The degree of similarity between the spectra from four of the mass spectrometers was calculated off-line by comparing the five most abundant ions from the spectrum on each instrument. A percentage fit value (20% for each ion that matches) was calculated by comparing each spectrum against the spectra recorded for the same compound on each instrument. The percentage of the inter-library pairwise comparisons (total = 434) that matched to greater than or equal to60% ranged from 64-89%, depending on the instrument pair. A blind trial was also undertaken using five unknown compounds resulting in 1670 pairwise comparisons with the library entries. The blind trial produced no false positives and correct identifications in all cases. The results of the study have established the basis for the construction of a transferable product-ion MS/MS library. Copyright (C) 2004 John Wiley Sons, Ltd.
引用
收藏
页码:1447 / 1454
页数:8
相关论文
共 13 条
[1]
EXPERIMENTAL INVESTIGATIONS OF FACTORS CONTROLLING THE COLLISION-INDUCED DISSOCIATION SPECTRA OF PEPTIDE IONS IN A TANDEM HYBRID MASS-SPECTROMETER .1. LEUCINE ENKEPHALIN [J].
ALEXANDER, AJ ;
BOYD, RK .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1989, 90 (03) :211-240
[2]
Baumann C, 2000, RAPID COMMUN MASS SP, V14, P349, DOI 10.1002/(SICI)1097-0231(20000315)14:5<349::AID-RCM873>3.3.CO
[3]
2-G
[4]
Evaluation of protocols for reproducible electrospray in-source collisionally induced dissociation on various liquid chromatography/mass spectrometry instruments and the development of spectral libraries [J].
Bristow, AWT ;
Nichols, WF ;
Webb, KS ;
Conway, B .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2002, 16 (24) :2374-2386
[5]
AN APPROXIMATE METHOD FOR DETERMINING THE EMITTANCE OF ION SOURCES IN QUADRUPOLE MASS-SPECTROMETRY [J].
DAWSON, PH ;
BINGQI, Y .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1983, 54 (1-2) :159-167
[6]
Lemire SW, 1996, J MASS SPECTROM, V31, P280, DOI 10.1002/(SICI)1096-9888(199603)31:3<280::AID-JMS296>3.0.CO
[7]
2-Y
[8]
Lopez LL, 1999, RAPID COMMUN MASS SP, V13, P663, DOI 10.1002/(SICI)1097-0231(19990430)13:8<663::AID-RCM538>3.0.CO
[9]
2-H
[10]
STANDARD REFERENCE SPECTRA FOR TANDEM MASS-SPECTROMETRY QUALITY ASSURANCE, PERFORMANCE EVALUATION AND PROFICIENCY TESTING - TANDEM MASS SPECTROMETERS WHICH USE RF-ONLY MULTIPOLE COLLISION CELLS [J].
MARTINEZ, RI .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1991, 5 (05) :245-248