Long-term biocompatibility of a corrodible peripheral iron stent in the porcine descending aorta

被引:499
作者
Peuster, Matthias [1 ]
Hesse, Carola
Schloo, Tirza
Fink, Christoph
Beerbaum, Philipp
von Schnakenburg, Christian
机构
[1] Univ Rostock, Dept Pediat Cardiol & Pediat Intens Care, Rostock, Germany
[2] German Pediat Heart Inst, Dept Pediat Thorac & Cardiovasc Surg, St Augustin, Germany
[3] Univ Freiburg, Dept Pediat & Adolescent Med, Freiburg, Germany
关键词
biodegradation; corrosion; degradation; intimal hyperplasia; stent;
D O I
10.1016/j.biomaterials.2006.05.029
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Currently there are no biodegradable stents available for treatment of vascular obstructions in patients with congenital heart defects. This study was performed to evaluate the safety of a corrodible stent produced from pure iron in a peripheral stent design (6-12 mm diameter) in a slotted tube design similar to a commercially available 316-L stent which served as control. Both stents were implanted into the descending aorta of 29 minipigs with an overstretch injury without technical problems. Two animals died after the implantation not related to the iron stent. The remaining 27 minipigs were followed for 1-360 days. Histomorphometry and quantitative angiography showed no difference with regard to the amount of neointimal proliferation between 316-L and iron stents. Histopathological examination of heart, lung, spleen, liver, kidney and para-aortic lymphatic nodes demonstrated no signs of iron overload or iron-related organ toxicity. Adjacent to the iron stent struts, there was no evidence for local toxicity due to corrosion products. We conclude that iron is a suitable metal for the production of a large-size degradable stent with no local or systemic toxicity. A faster degradation rate, however, is desirable and further studies have to focus on the modification of the composition and design of the stent to expedite the degradation process. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:4955 / 4962
页数:8
相关论文
共 42 条
[1]   Molecular control of iron metabolism [J].
Andrews, NC .
BEST PRACTICE & RESEARCH CLINICAL HAEMATOLOGY, 2005, 18 (02) :159-169
[2]  
BAURSCHMIDT P, 1977, J BIOENG, V1, P261
[3]  
Di Mario Carlo, 2004, J Interv Cardiol, V17, P391, DOI 10.1111/j.1540-8183.2004.04081.x
[4]   A METHOD FOR THE STUDY OF UNDECALCIFIED BONES AND TEETH WITH ATTACHED SOFT-TISSUES - THE SAGE-SCHLIFF (SAWING AND GRINDING) TECHNIQUE [J].
DONATH, K ;
BREUNER, G .
JOURNAL OF ORAL PATHOLOGY & MEDICINE, 1982, 11 (04) :318-326
[5]   TRANSLUMINALLY-PLACED COILSPRING ENDARTERIAL TUBE GRAFTS - LONG-TERM PATENCY IN CANINE POPLITEAL ARTERY [J].
DOTTER, CT .
INVESTIGATIVE RADIOLOGY, 1969, 4 (05) :329-+
[6]   Morphological predictors of restenosis after coronary stenting in humans [J].
Farb, A ;
Weber, DK ;
Kolodgie, FD ;
Burke, AP ;
Virmani, R .
CIRCULATION, 2002, 105 (25) :2974-2980
[7]   Coronary stenting and inflammation [J].
Gaspardone, A ;
Versaci, F .
AMERICAN JOURNAL OF CARDIOLOGY, 2005, 96 (12A) :65L-70L
[8]   Acute and chronic tissue response to coronary stent implantation:: Pathologic findings in human specimen [J].
Grewe, PH ;
Deneke, T ;
Machraoui, A ;
Barmeyer, J ;
Müller, KM .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (01) :157-163
[9]   Iron homeostasis and inherited iron overload disorders: an overview [J].
Heeney, MM ;
Andrews, NC .
HEMATOLOGY-ONCOLOGY CLINICS OF NORTH AMERICA, 2004, 18 (06) :1379-+
[10]  