Validation of the Omron 705 IT oscillometric device for home blood pressure measurement in children and adolescents: The Arsakion School Study

被引:147
作者
Stergiou, George S. [1 ]
Yiannes, Nikolaos G. [1 ]
Rarra, Vayia C. [1 ]
机构
[1] Sotiria Hosp, Univ Dept Med 3, Hypertens Ctr, Athens 11527, Greece
关键词
accuracy; European Society of Hypertension; home blood pressure; international protocol; Omron; self-measurement; validation;
D O I
10.1097/01.mbp.0000209074.38331.16
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background Limited evidence exists on the accuracy of oscillometric devices for blood pressure measurement in children. This study validated the Omron 705 IT monitor (Omron Healthcare Europe BV, Hoofddorp, The Netherlands) in normotensive children and adolescents. Methods Simultaneous blood pressure measurements were taken by two observers (connected mercury sphygmomanometers) four times, sequentially with three measurements by using the tested device. Absolute device-observer blood pressure differences were classified into three zones (within 5, 10 and 15 mmHg) and assessed using the European Society of Hypertension International Protocol criteria. The number of readings with a difference within 5 mmHg was calculated for each individual. The American Association for the Advancement of Medical Instrumentation criterion (mean difference +/- SD <5 +/- 8 mmHg) was also applied. Results A total of 197 study participants (591 readings) were included (99 boys, mean age +/- SD 10.6 +/- 2.4 years (range 6-16 years). The device produced 378 (64%), 532 (90%) and 579 (98%) measurements within 5, 10 and 15 mmHg, respectively, for systolic blood pressure, and 415 (70%), 525 (89%) and 577 (98%) measurements, respectively, for diastolic blood pressure. In all, 136 study participants (69%) had at least two systolic blood pressure differences within 5 mmHg and 16 (8%) had no differences within 5 mmHg [for diastolic blood pressure 147 (75%) and nine (5%) participants, respectively]. Mean systolic blood pressure difference was 4.0 +/- 4.8 mmHg and diastolic blood pressure -2.1 +/- 5.9 mmHg and was <5 +/- 8 mmHg in quartiles of participants divided by age, height, body mass index, arm circumference, blood pressure or pulse rate. Conclusions The Omron 705 IT appears to be an accurate device for blood pressure measurement in normotensive children and adolescents.
引用
收藏
页码:229 / 234
页数:6
相关论文
共 17 条
  • [1] Association for the Advancement of Medical Instrumentation, 1993, AM NAT STAND EL AUT
  • [2] Measurement of blood pressure at home: survey among pediatric nephrologists
    Bald, M
    Hoyer, PF
    [J]. PEDIATRIC NEPHROLOGY, 2001, 16 (12) : 1058 - 1062
  • [3] Belsha, 1996, Blood Press Monit, V1, P127
  • [4] Center for Disease Control and Prevention, 2000 CDC GROWTH CHAR
  • [5] DABL Educational Trust, DEV BLOOD PRESS MEAS
  • [6] Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT
    El Assaad, MA
    Topouchian, JA
    Asmar, RG
    [J]. BLOOD PRESSURE MONITORING, 2003, 8 (03) : 127 - 133
  • [7] Evaluation of the Omron 705-CP blood pressure monitor for use in children and adolescents
    Furusawa, EA
    Ruiz, MFO
    Saito, MI
    Koch, VH
    [J]. JOURNAL OF HYPERTENSION, 2004, 22 : S140 - S140
  • [8] Modesti, 1996, Blood Press Monit, V1, P63
  • [9] Trends in blood pressure among children and adolescents
    Muntner, P
    He, J
    Cutler, JA
    Wildman, RP
    Whelton, PK
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2004, 291 (17): : 2107 - 2113
  • [10] Reporting bias in self-measurement of blood pressure
    Myers, MG
    [J]. BLOOD PRESSURE MONITORING, 2001, 6 (04) : 181 - 183