Duplex ultrasound criteria for the identification of carotid stenosis should be laboratory specific

被引:72
作者
Kuntz, KM
Polak, JF
Whittemore, AD
Skillman, JJ
Kent, KC
机构
[1] HARVARD UNIV,BETH ISRAEL HOSP,SCH MED,DEPT SURG,BOSTON,MA 02215
[2] BRIGHAM & WOMENS HOSP,DEPT SURG,BOSTON,MA 02115
[3] BRIGHAM & WOMENS HOSP,DEPT RADIOL,BOSTON,MA 02115
[4] BRIGHAM & WOMENS HOSP,DEPT MED,BOSTON,MA 02115
关键词
carotid stenosis; diagnosis; ultrasonics;
D O I
10.1161/01.STR.28.3.597
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background and Purpose Published criteria for the determination of carotid stenosis have been widely applied by vascular laboratories. We compared two vascular laboratories and their duplex ultrasound (DU) machines in terms of their overall diagnostic performance and the optimal criteria to identify patients who have a 70% to 99% stenosis of the internal carotid artery. Methods Measurements of stenosis by DU and angiography were compared for 123 carotid arteries (60 arteries, laboratory A; 63 arteries, laboratory B). Receiver operating characteristic (ROC) curves were created, and the areas under the ROC curves and the optimal criteria for determining a 70% to 99% stenosis were compared. Multiple regression analysis was used to measure the effect of laboratory on the relationship between angiographic stenosis and DU velocity parameters. Results Areas under the ROC curves were similar for both laboratories (0.89 to 0.90, laboratory A; 0.90 to 0.92, laboratory B). However, the optimal criterion for the identification of a 70% to 99% carotid stenosis was different for each laboratory. For most velocity parameters, based on regression analyses, the predicted percent angiographic stenosis for laboratory A was significantly greater than that for laboratory B. In addition, performance differed between the laboratories when established criteria from the literature were applied. Conclusions Two vascular laboratories with similar diagnostic accuracy by ROC analysis have markedly different ''optimal'' DU criteria. For a given angiographic stenosis, velocities in one laboratory were consistently greater than those in the other laboratory. Laboratory-specific criteria rather than published criteria should be used to identify patients with internal carotid artery stenoses.
引用
收藏
页码:597 / 602
页数:6
相关论文
共 21 条
  • [1] Chervu A, 1994, Ann Vasc Surg, V8, P296, DOI 10.1007/BF02018179
  • [2] Daigle RJ, 1990, J VASC TECHNOL, V14, P206
  • [3] THE ROLE OF DUPLEX SCANNING AND ARTERIOGRAPHY BEFORE CAROTID ENDARTERECTOMY - A PROSPECTIVE-STUDY
    DAWSON, DL
    ZIERLER, RE
    STRANDNESS, DE
    CLOWES, AW
    KOHLER, TR
    [J]. JOURNAL OF VASCULAR SURGERY, 1993, 18 (04) : 673 - 683
  • [4] 1977 RIETZ LECTURE - BOOTSTRAP METHODS - ANOTHER LOOK AT THE JACKKNIFE
    EFRON, B
    [J]. ANNALS OF STATISTICS, 1979, 7 (01) : 1 - 26
  • [5] *EX COMM AS CAR AT, 1995, JAMA-J AM MED ASSOC, V273, P1421
  • [6] COLOR-FLOW DUPLEX SCANNING OF CAROTID ARTERIES - NEW VELOCITY CRITERIA-BASED ON RECEIVER OPERATOR CHARACTERISTIC ANALYSIS FOR THRESHOLD STENOSES USED IN THE SYMPTOMATIC AND ASYMPTOMATIC CAROTID TRIALS
    FAUGHT, WE
    MATTOS, MA
    VANBEMMELEN, PS
    HODGSON, KJ
    BARKMEIER, LD
    RAMSEY, DE
    SUMNER, DS
    [J]. JOURNAL OF VASCULAR SURGERY, 1994, 19 (05) : 818 - 828
  • [7] IS DUPLEX SCANNING SUFFICIENT EVALUATION BEFORE CAROTID ENDARTERECTOMY
    GEUDER, JW
    LAMPARELLO, PJ
    RILES, TS
    GIANGOLA, G
    IMPARATO, AM
    [J]. JOURNAL OF VASCULAR SURGERY, 1989, 9 (02) : 193 - 201
  • [8] CAN CAROTID DUPLEX SCANNING SUPPLANT ARTERIOGRAPHY IN PATIENTS WITH FOCAL CAROTID TERRITORY SYMPTOMS
    GOODSON, SF
    FLANIGAN, DP
    BISHARA, RA
    SCHULER, JJ
    KIKTA, MJ
    MEYER, JP
    [J]. JOURNAL OF VASCULAR SURGERY, 1987, 5 (04) : 551 - 557
  • [9] THE MEANING AND USE OF THE AREA UNDER A RECEIVER OPERATING CHARACTERISTIC (ROC) CURVE
    HANLEY, JA
    MCNEIL, BJ
    [J]. RADIOLOGY, 1982, 143 (01) : 29 - 36
  • [10] Howard G, 1991, J Stroke Cerebrovasc Dis, V1, P166, DOI 10.1016/S1052-3057(10)80013-8