Spectrometric system for characterizing drop and powder trajectories and chemistry in reactive flows

被引:7
作者
Cabalo, JC
Schmidt, J
Wendt, JOL
Scheeline, A
机构
[1] Univ Illinois, Dept Chem, Urbana, IL 61801 USA
[2] Univ Arizona, Dept Chem & Environm Engn, Tucson, AZ 85721 USA
关键词
flame diagnostics; planar laser-induced fluorescence; spectroscopy in harsh environments; imaging; drop vaporization;
D O I
10.1366/000370202760355073
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An instrument for emission, fluorescence, and scattering diagnostics of a 22 kW gas/air thermal reactor is described. Z-pair parabolic mirrors provide diffraction-limited imaging in field center, with worst-case blur over a 12.8 mm field of view having a radius of 0.044 mm rms. Either full-field filtered or transverse line dispersed viewing can occur simultaneously. Dispersion is achieved using a 2-m spectrograph f-matched to the feed optics. Examples of data show non-axisymmetric flow in a nominally laminar flow system, drop shapes in situ, and Mie scattering from soot.
引用
收藏
页码:1345 / 1353
页数:9
相关论文
共 39 条
[31]   Time-resolved measurements with single droplet introduction to investigate space-charge effects in plasma mass spectrometry [J].
Stewart, II ;
Olesik, JW .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1999, 10 (02) :159-174
[32]  
STEWART TL, 1997, 214 ACS NAT M LAS VE
[33]  
TERRELL CW, 1994, NUCL MAT MANAGE, V23, P233
[34]  
TOLLISON SA, 1997, RAD PROT MANAGE, V14, P45
[35]  
WAINNER RT, 1999, THESIS GEORGIA I TEC
[36]   Time-resolved laser-induced incandescence and laser elastic-scattering measurements in a propane diffusion flame [J].
Witze, PO ;
Hochgreb, S ;
Kayes, D ;
Michelsen, HA ;
Shaddix, CR .
APPLIED OPTICS, 2001, 40 (15) :2443-2452
[37]  
ZHANG JB, 1991, J OPT SOC AM B, V8, P319
[38]  
ZOELLNER MJ, 1987, APPL SPECTROSC, V91, P943
[39]   Hanford's nuclear wasteland [J].
Zorpette, G .