Validated quantitation method for a peptide in rat serum using liquid chromatography/high-field asymmetric waveform ion mobility spectrometry

被引:50
作者
Klaassen, Tobias [1 ]
Szwandt, Simon [2 ]
Kapron, James T. [3 ]
Roemer, Axel [1 ]
机构
[1] Anal & Metab Forsch Serv GmbH, A&M Lab, D-50126 Bergheim, Germany
[2] Thermo Fisher Sci, Hemel Hempstead HP2 7GE, England
[3] Thermo Fisher Sci, Nepean, ON K2E 7L6, Canada
关键词
MASS-SPECTROMETRY; FAIMS-MS; QUANTIFICATION; PROTEOMICS;
D O I
10.1002/rcm.4147
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The analysis of peptides presents serious challenges for bioanalytical scientists including low total ion current and non-selective fragmentation during tandem mass spectrometry (MS/MS). During method validation of a peptide in rat serum matrix some interferences could not be easily removed and thus prevented accurate and precise measurement. These problems associated with peptide quantitation were resolved by using FAIMS (high-Field Asymmetric waveform Ion Mobility Spectrometry). This selectivity-enhancing technique filters out matrix interferences, and the resulting pseudo-selected reaction monitoring (pseudo-SRM) chromatograms were nearly free from interferences. Control blank matrix samples contained an acceptable level of interference (only 7% signal as compared to the lower level of quantitation). Chromatographic peaks were easily, accurately and precisely integrated resulting in a validated liquid chromatography (LC)/FAIMS-MS/MS method for the analysis of a peptide drug in rat serum according to United States Food and Drug Administration (US FDA) bioanalytical guidelines. These results confirm that new selectivity-enhancing technologies aid the pharmaceutical industry in reliably producing acceptable pharmacokinetic data. Copyright (C) 2009 John Wiley & Sons, Ltd.
引用
收藏
页码:2301 / 2306
页数:6
相关论文
共 12 条
[1]   Mass spectrometry-based proteomics [J].
Aebersold, R ;
Mann, M .
NATURE, 2003, 422 (6928) :198-207
[2]  
[Anonymous], 2001, Guidance for Industry Bioanalytical Method Validation
[3]   Atmospheric pressure ionization LC/MS/MS techniques for drug disposition studies [J].
Brewer, E ;
Henion, J .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1998, 87 (04) :395-402
[4]   Assessing the dynamic range and peak capacity of nanoflow LC-FAIMS-MS on an ion trap mass spectrometer for proteomics [J].
Canterbury, Jesse D. ;
Yi, Xianhua ;
Hoopmann, Michael R. ;
MacCoss, Michael J. .
ANALYTICAL CHEMISTRY, 2008, 80 (18) :6888-6897
[5]   Trace level impurity method development with high-field asymmetric waveform ion mobility spectrometry: systematic study of factors affecting the performance [J].
Champarnaud, Elodle ;
Laures, Alice M. -F. ;
Borman, Phil J. ;
Chatfield, Marion J. ;
Kapron, James T. ;
Harrison, Mark ;
Wolff, Jean-Claude .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2009, 23 (01) :181-193
[6]  
Hatsis P, 2008, RAPID COMMUN MASS SP, V22, P735, DOI 10.1002/rcm.3416
[7]   Ion mobility-mass spectrometry [J].
Kanu, Abu B. ;
Dwivedi, Prabha ;
Tam, Maggie ;
Matz, Laura ;
Hill, Herbert H., Jr. .
JOURNAL OF MASS SPECTROMETRY, 2008, 43 (01) :1-22
[8]  
Lee MS, 1999, MASS SPECTROM REV, V18, P187, DOI 10.1002/(SICI)1098-2787(1999)18:3/4<187::AID-MAS2>3.0.CO
[9]  
2-K
[10]   Terbutaline enantiomer separation and quantification by complexation and field asymmetric ion mobility spectrometry-tandem mass spectrometry [J].
Mie, Axel ;
Ray, Andrew ;
Axelsson, Bengt-Olof ;
Jornten-Karlsson, Magnus ;
Reimann, Curt T. .
ANALYTICAL CHEMISTRY, 2008, 80 (11) :4133-4140