Trace level impurity method development with high-field asymmetric waveform ion mobility spectrometry: systematic study of factors affecting the performance

被引:11
作者
Champarnaud, Elodle [1 ]
Laures, Alice M. -F. [1 ]
Borman, Phil J. [1 ]
Chatfield, Marion J. [2 ]
Kapron, James T. [3 ]
Harrison, Mark [4 ]
Wolff, Jean-Claude [1 ]
机构
[1] Glaxo SmithKline, Analyt Sci, Stevenage SG1 2NY, Herts, England
[2] GlaxoSmithKline Inc, Stat Sci, Harlow CM19 5AW, Essex, England
[3] Thermo Fisher Sci, Nepean, ON K2E 7L6, Canada
[4] Thermo Fisher Sci, Hemel Hempstead HP2 7GE, England
关键词
MASS-SPECTROMETRY; FAIMS; SEPARATION; DESIGN; TOOL;
D O I
10.1002/rcm.3844
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
For the determination of trace level impurities, analytical chemists are confronted with complex mixtures and difficult separations. New technologies such as high-field asymmetric waveform ion mobility spectrometry (FAIMS) have been developed to make their work easier; however, efficient method development and troubleshooting can be quite challenging if little prior knowledge of the factors or their settings is available. We present the results of an investigation performed in order to obtain a better understanding of the FAIMS technology. The influence of eight factors (polarity of dispersion voltage, outer bias voltage, total gas flow rate, composition of the carrier gas (e.g. %He), outer electrode temperature, ratio between the temperatures, of the inner and outer electrodes, flow rate and composition of the make-up mobile phase) was assessed. Five types of responses were monitored: value of the compensation voltage (CV), intensity, width and asymmetry of the compensation voltage peak, and resolution between two peaks. Three types of studies were performed using different test mixtures and various ionisation modes to assess whether the same conclusions could be drawn across these conditions for a number of different types of compounds. To extract the maximum information from as few experiments as possible, a Design of Experiment (DoE) approach was used. The results presented in this work provide detailed information on the factors affecting FAIMS separations and therefore should enable the user to troubleshoot more effectively and to develop efficient methods. Copyright (C) 2008 John Wiley & Sons, Ltd.
引用
收藏
页码:181 / 193
页数:13
相关论文
共 11 条
[1]   Characterization of a temperature-controlled FAIMS system [J].
Barnett, David A. ;
Belford, Michael ;
Dunyach, Jean-Jacques ;
Purves, Randy W. .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2007, 18 (09) :1653-1663
[2]   Separation of β2-microglobulin conformers by high-field asymmetric waveform ion mobility spectrometry (FAIMS) coupled to electrospray ionisation mass spectrometry [J].
Borysik, AJH ;
Read, P ;
Little, DR ;
Bateman, RH ;
Radford, SE ;
Ashcroft, AE .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (19) :2229-2234
[3]   Evaluating the utility of ion mobility separation in combination with high-pressure liquid chromatography/mass spectrometry to facilitate detection of trace impurities in formulated drug products [J].
Eckers, Christine ;
Laures, Alice M. -F. ;
Giles, Kevin ;
Major, Hilary ;
Pringle, Steve .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2007, 21 (07) :1255-1263
[4]   High-field asymmetric waveform ion mobility spectrometry: A new tool for mass spectrometry [J].
Guevremont, R .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1058 (1-2) :3-19
[5]  
Hatsis P, 2008, RAPID COMMUN MASS SP, V22, P735, DOI 10.1002/rcm.3416
[6]   Simultaneous analysis of prostanoids using liquid chromatography/high-field asymmetric waveform ion mobility spectrometry/tandem mass spectrometry [J].
Kapron, James ;
Wu, Jin ;
Mauriala, Timo ;
Clark, Patsy ;
Purves, Randy W. ;
Bateman, Kevin P. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2006, 20 (10) :1504-1510
[7]   Review of applications of high-field asymmetric waveform ion mobility spectrometry (FAIMS) and differential mobility spectrometry (DMS) [J].
Kolakowski, Beata M. ;
Mester, Zoltan .
ANALYST, 2007, 132 (09) :842-864
[8]   Investigation into the factors affecting accuracy of mass measurements on a time-of-flight mass spectrometer using Design of Experiment [J].
Laures, Alice M. -F. ;
Wolff, Jean-Claude ;
Eckers, Christine ;
Borman, Phil J. ;
Chatfield, Marion J. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2007, 21 (04) :529-535
[9]   Statistical design of experiments as a tool in mass spectrometry [J].
Riter, LS ;
Vitek, O ;
Gooding, KM ;
Hodge, BD ;
Julian, RK .
JOURNAL OF MASS SPECTROMETRY, 2005, 40 (05) :565-579
[10]   High-field asymmetric waveform ion mobility spectrometry coupled with liquid chromatography/electrospray ionization tandem mass spectrometry (LOESI-FAIMS-MSMS) multi-component bioanalytical method development, performance evaluation and demonstration of the constancy of the compensation voltage with change of mobile phase composition or flow rate [J].
Wu, Steven T. ;
Xia, Yuan-Qing ;
Jemal, Mohammed .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2007, 21 (22) :3667-3676