Outpatient Glycemic Control with a Bionic Pancreas in Type 1 Diabetes

被引:409
作者
Russell, Steven J. [1 ,2 ,3 ]
El-Khatib, Firas H. [4 ]
Sinha, Manasi [2 ,3 ]
Magyar, Kendra L. [1 ,2 ,3 ]
McKeon, Katherine [4 ]
Goergen, Laura G. [1 ,2 ,3 ]
Balliro, Courtney [1 ,2 ,3 ]
Hillard, Mallory A. [1 ,2 ,3 ]
Nathan, David M. [1 ,2 ,3 ]
Damiano, Edward R. [4 ]
机构
[1] Massachusetts Gen Hosp, Diabet Unit, Boston, MA 02114 USA
[2] Massachusetts Gen Hosp, Dept Med, Boston, MA 02114 USA
[3] Harvard Univ, Sch Med, Boston, MA USA
[4] Boston Univ, Dept Biomed Engn, Boston, MA 02215 USA
关键词
LOOP INSULIN DELIVERY; ARTIFICIAL PANCREAS; SEVERE HYPOGLYCEMIA; GLUCOSE CONTROL; ADOLESCENTS; ADULTS; KETOACIDOSIS; FEASIBILITY; SYSTEM; YOUTH;
D O I
10.1056/NEJMoa1314474
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND The safety and effectiveness of automated glycemic management have not been tested in multiday studies under unrestricted outpatient conditions. METHODS In two random-order, crossover studies with similar but distinct designs, we compared glycemic control with a wearable, bihormonal, automated, "bionic" pancreas (bionic-pancreas period) with glycemic control with an insulin pump (control period) for 5 days in 20 adults and 32 adolescents with type 1 diabetes mellitus. The automatically adaptive algorithm of the bionic pancreas received data from a continuous glucose monitor to control subcutaneous delivery of insulin and glucagon. RESULTS Among the adults, the mean plasma glucose level over the 5-day bionic-pancreas period was 138 mg per deciliter (7.7 mmol per liter), and the mean percentage of time with a low glucose level (<70 mg per deciliter [3.9 mmol per liter]) was 4.8%. After 1 day of automatic adaptation by the bionic pancreas, the mean (+/- SD) glucose level on continuous monitoring was lower than the mean level during the control period (133 +/- 13 vs. 159 +/- 30 mg per deciliter [7.4 +/- 0.7 vs. 8.8 +/- 1.7 mmol per liter], P<0.001) and the percentage of time with a low glucose reading was lower (4.1% vs. 7.3%, P = 0.01). Among the adolescents, the mean plasma glucose level was also lower during the bionic-pancreas period than during the control period (138 +/- 18 vs. 157 +/- 27 mg per deciliter [7.7 +/- 1.0 vs. 8.7 +/- 1.5 mmol per liter], P = 0.004), but the percentage of time with a low plasma glucose reading was similar during the two periods (6.1% and 7.6%, respectively; P = 0.23). The mean frequency of interventions for hypoglycemia among the adolescents was lower during the bionic-pancreas period than during the control period (one per 1.6 days vs. one per 0.8 days, P<0.001). CONCLUSIONS As compared with an insulin pump, a wearable, automated, bihormonal, bionic pancreas improved mean glycemic levels, with less frequent hypoglycemic episodes, among both adults and adolescents with type 1 diabetes mellitus.
引用
收藏
页码:313 / 325
页数:13
相关论文
共 37 条
[1]  
Amer Diabet Assoc, 2011, DIABETES CARE, V34, pS11, DOI [10.2337/dc10-S062, 10.2337/dc14-S081, 10.2337/dc11-S011, 10.2337/dc13-S011, 10.2337/dc13-S067, 10.2337/dc12-s064, 10.2337/dc11-S062, 10.2337/dc10-S011, 10.2337/dc12-s011]
[2]   Hypoglycemia in the diabetes control and complications trial [J].
不详 .
DIABETES, 1997, 46 (02) :271-286
[3]  
[Anonymous], 2009, DIABETES CARE
[4]  
[Anonymous], DEXCOM G4 PLATINUM C
[5]   MD-Logic Artificial Pancreas System A pilot study in adults with type 1 diabetes [J].
Atlas, Eran ;
Nimri, Revital ;
Miller, Shahar ;
Grunberg, Eli A. ;
Phillip, Moshe .
DIABETES CARE, 2010, 33 (05) :1072-1076
[6]   The T1D Exchange Clinic Registry [J].
Beck, Roy W. ;
Tamborlane, William V. ;
Bergenstal, Richard M. ;
Miller, Kellee M. ;
DuBose, Stephanie N. ;
Hall, Callyn A. .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2012, 97 (12) :4383-4389
[7]   Missed insulin meal boluses and elevated hemoglobin A1c levels in children receiving insulin pump therapy [J].
Burdick, J ;
Chase, HP ;
Slover, RH ;
Knievel, K ;
Scrimgeour, L ;
Maniatis, AK ;
Klingensmith, GJ .
PEDIATRICS, 2004, 113 (03) :E221-E224
[8]   Novel Use of Glucagon in a Closed-Loop System for Prevention of Hypoglycemia in Type 1 Diabetes [J].
Castle, Jessica R. ;
Engle, Julia M. ;
Youssef, Joseph E. L. ;
Massoud, Ryan G. ;
Yuen, Kevin C. J. ;
Kagan, Ryland ;
Ward, W. Kenneth .
DIABETES CARE, 2010, 33 (06) :1282-1287
[9]   Severe hypoglycemia and diabetic ketoacidosis among youth with type 1 diabetes in the T1D Exchange clinic registry [J].
Cengiz, Eda ;
Xing, Dongyuan ;
Wong, Jenise C. ;
Wolfsdorf, Joseph I. ;
Haymond, Morey W. ;
Rewers, Arleta ;
Shanmugham, Satya ;
Tamborlane, William V. ;
Willi, Steven M. ;
Seiple, Diane L. ;
Miller, Kellee M. ;
DuBose, Stephanie N. ;
Beck, Roy W. .
PEDIATRIC DIABETES, 2013, 14 (06) :447-454
[10]   Hypoglycemia in diabetes [J].
Cryer, PE ;
Davis, SN ;
Shamoon, H .
DIABETES CARE, 2003, 26 (06) :1902-1912