Outpatient Glycemic Control with a Bionic Pancreas in Type 1 Diabetes

被引:409
作者
Russell, Steven J. [1 ,2 ,3 ]
El-Khatib, Firas H. [4 ]
Sinha, Manasi [2 ,3 ]
Magyar, Kendra L. [1 ,2 ,3 ]
McKeon, Katherine [4 ]
Goergen, Laura G. [1 ,2 ,3 ]
Balliro, Courtney [1 ,2 ,3 ]
Hillard, Mallory A. [1 ,2 ,3 ]
Nathan, David M. [1 ,2 ,3 ]
Damiano, Edward R. [4 ]
机构
[1] Massachusetts Gen Hosp, Diabet Unit, Boston, MA 02114 USA
[2] Massachusetts Gen Hosp, Dept Med, Boston, MA 02114 USA
[3] Harvard Univ, Sch Med, Boston, MA USA
[4] Boston Univ, Dept Biomed Engn, Boston, MA 02215 USA
关键词
LOOP INSULIN DELIVERY; ARTIFICIAL PANCREAS; SEVERE HYPOGLYCEMIA; GLUCOSE CONTROL; ADOLESCENTS; ADULTS; KETOACIDOSIS; FEASIBILITY; SYSTEM; YOUTH;
D O I
10.1056/NEJMoa1314474
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND The safety and effectiveness of automated glycemic management have not been tested in multiday studies under unrestricted outpatient conditions. METHODS In two random-order, crossover studies with similar but distinct designs, we compared glycemic control with a wearable, bihormonal, automated, "bionic" pancreas (bionic-pancreas period) with glycemic control with an insulin pump (control period) for 5 days in 20 adults and 32 adolescents with type 1 diabetes mellitus. The automatically adaptive algorithm of the bionic pancreas received data from a continuous glucose monitor to control subcutaneous delivery of insulin and glucagon. RESULTS Among the adults, the mean plasma glucose level over the 5-day bionic-pancreas period was 138 mg per deciliter (7.7 mmol per liter), and the mean percentage of time with a low glucose level (<70 mg per deciliter [3.9 mmol per liter]) was 4.8%. After 1 day of automatic adaptation by the bionic pancreas, the mean (+/- SD) glucose level on continuous monitoring was lower than the mean level during the control period (133 +/- 13 vs. 159 +/- 30 mg per deciliter [7.4 +/- 0.7 vs. 8.8 +/- 1.7 mmol per liter], P<0.001) and the percentage of time with a low glucose reading was lower (4.1% vs. 7.3%, P = 0.01). Among the adolescents, the mean plasma glucose level was also lower during the bionic-pancreas period than during the control period (138 +/- 18 vs. 157 +/- 27 mg per deciliter [7.7 +/- 1.0 vs. 8.7 +/- 1.5 mmol per liter], P = 0.004), but the percentage of time with a low plasma glucose reading was similar during the two periods (6.1% and 7.6%, respectively; P = 0.23). The mean frequency of interventions for hypoglycemia among the adolescents was lower during the bionic-pancreas period than during the control period (one per 1.6 days vs. one per 0.8 days, P<0.001). CONCLUSIONS As compared with an insulin pump, a wearable, automated, bihormonal, bionic pancreas improved mean glycemic levels, with less frequent hypoglycemic episodes, among both adults and adolescents with type 1 diabetes mellitus.
引用
收藏
页码:313 / 325
页数:13
相关论文
共 37 条
[11]  
El Youssef Joseph, 2011, J Diabetes Sci Technol, V5, P1312
[12]   Autonomous and Continuous Adaptation of a Bihormonal Bionic Pancreas in Adults and Adolescents With Type 1 Diabetes [J].
El-Khatib, Firas H. ;
Russell, Steven J. ;
Magyar, Kendra L. ;
Sinha, Manasi ;
McKeon, Katherine ;
Nathan, David M. ;
Damiano, Edward R. .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2014, 99 (05) :1701-1711
[13]   A Bihormonal Closed-Loop Artificial Pancreas for Type 1 Diabetes [J].
El-Khatib, Firas H. ;
Russell, Steven J. ;
Nathan, David M. ;
Sutherlin, Robert G. ;
Damiano, Edward R. .
SCIENCE TRANSLATIONAL MEDICINE, 2010, 2 (27) :27ra27
[14]   Closed-Loop Basal Insulin Delivery Over 36 Hours in Adolescents With Type 1 Diabetes [J].
Elleri, Daniela ;
Allen, Janet M. ;
Kumareswaran, Kavita ;
Leelarathna, Lalantha ;
Nodale, Marianna ;
Caldwell, Karen ;
Cheng, Peiyao ;
Kollman, Craig ;
Haidar, Ahmad ;
Murphy, Helen R. ;
Wilinska, Malgorzata E. ;
Acerini, Carlo L. ;
Dunger, David B. ;
Hovorka, Roman .
DIABETES CARE, 2013, 36 (04) :838-844
[15]  
Genuth S, 2002, JAMA-J AM MED ASSOC, V287, P2563
[16]   Overnight Closed- Loop Insulin Delivery in Young People With Type 1 Diabetes: A Free- Living, Randomized Clinical Trial [J].
Hovorka, Roman ;
Elleri, Daniela ;
Thabit, Hood ;
Allen, Janet M. ;
Leelarathna, Lalantha ;
El-Khairi, Ranna ;
Kumareswaran, Kavita ;
Caldwell, Karen ;
Calhoun, Peter ;
Kollman, Craig ;
Murphy, Helen R. ;
Acerini, Carlo L. ;
Wilinska, Malgorzata E. ;
Nodale, Marianna ;
Dunger, David B. .
DIABETES CARE, 2014, 37 (05) :1204-1211
[17]   Feasibility of Outpatient Fully Integrated Closed-Loop Control First studies of wearable artificial pancreas [J].
Kovatchev, Boris P. ;
Renard, Eric ;
Cobelli, Claudio ;
Zisser, Howard C. ;
Keith-Hynes, Patrick ;
Anderson, Stacey M. ;
Brown, Sue A. ;
Chernavvsky, Daniel R. ;
Breton, Marc D. ;
Farret, Anne ;
Pelletier, Marie-Josee ;
Place, Jerome ;
Bruttomesso, Daniela ;
Del Favero, Simone ;
Visentin, Roberto ;
Filippi, Alessio ;
Scotton, Rachele ;
Avogaro, Angelo ;
Doyle, Francis J., III .
DIABETES CARE, 2013, 36 (07) :1851-1858
[18]   Closed- Loop Artificial Pancreas Systems: Physiological Input to Enhance Next- Generation Devices [J].
Kudva, Yogish C. ;
Carter, Rickey E. ;
Cobelli, Claudio ;
Basu, Rita ;
Basu, Ananda .
DIABETES CARE, 2014, 37 (05) :1184-1190
[19]   Frequency of severe hypoglycemia requiring emergency treatment in type 1 and type 2 diabetes - A population-based study of health service resource use [J].
Leese, GP ;
Wang, JX ;
Broomhall, J ;
Kelly, P ;
Marsden, A ;
Morrison, W ;
Frier, BM ;
Morris, AD .
DIABETES CARE, 2003, 26 (04) :1176-1180
[20]   DAY-TO-DAY VARIATION OF CONTINUOUSLY MONITORED GLYCEMIA - FURTHER MEASURE OF DIABETIC INSTABILITY [J].
MOLNAR, GD ;
TAYLOR, WF ;
HO, MM .
DIABETOLOGIA, 1972, 8 (05) :342-&