Integrating microfabricated fluidic systems and NMR spectroscopy

被引:89
作者
Trumbull, JD
Glasgow, IK
Beebe, DJ
Magin, RL
机构
[1] Univ Illinois, Beckman Inst Adv Sci & Technol, Urbana, IL 61801 USA
[2] Univ Illinois, Dept Elect & Comp Engn, Urbana, IL 61801 USA
基金
美国国家科学基金会;
关键词
capillary electrophoresis (CE); microcoil; microfluidic; nuclear magnetic resonance spectroscopy (NMR); total analysis system;
D O I
10.1109/10.817611
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The philosophy of miniature total analysis systems (mu-TAS) hinges on the integration of multiple chemical processing steps and the means of analyzing their results on the same miniaturized system. We have constructed chip-based capillary electrophoresis (CE) devices equipped with an integrated planar radio-frequency detector coil used for nuclear magnetic resonance spectroscopy (NMR). Separations were accomplished in the devices, but satisfactory NMR spectra could only be obtained from samples of high concentration. The relative sensitivity is explained and the scaling law dichotomy of CE and NMR explored.
引用
收藏
页码:3 / 7
页数:5
相关论文