Validation of a telemetry system for long-term measurement of blood pressure

被引:33
作者
Brooks, D [1 ]
Horner, RL [1 ]
Kozar, LF [1 ]
Waddell, TK [1 ]
Render, CL [1 ]
Phillipson, EA [1 ]
机构
[1] UNIV TORONTO,DEPT MED,TORONTO,ON M5S 1A8,CANADA
关键词
radiotelemetry; chronic monitoring; cardiovascular system;
D O I
10.1152/jappl.1996.81.2.1012
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
We have used an implanted telemetry system to continuously monitor blood pressure (BP) in three dogs for durations ranging from 28 to 75 wk after implantation. Measurements of BP obtained by telemetry were compared every 3-12 wk, with measurements of BP recorded with a manometer-tipped catheter that was inserted into a femoral artery. Over a wide range of both physiological and pharmacologically manipulated pressures (40-200 mmHg), the values of BP obtained by the two methods were highly correlated (all r >0.966; all P less than or equal to 0.0001). However, the mean differences between the values obtained by the two systems were different from zero (range +29.6 to -1.5 mmHg; P less than or equal to 0.0001), indicating an offset in the BP values recorded from the implanted system. Furthermore, this offset was dependent on the absolute level of the BP. The findings indicate that, for a period of at least 28 wk and up to 75 wk after implantation, the telemetry system accurately measures acute changes in BP and can reliably measure absolute BP provided that the system is properly validated.
引用
收藏
页码:1012 / 1018
页数:7