Catheter visualization using locally induced, actively controlled field inhomogeneities

被引:94
作者
Glowinski, A [1 ]
Adam, G [1 ]
Bucker, A [1 ]
Neuerburg, J [1 ]
vanVaals, JJ [1 ]
Gunther, RW [1 ]
机构
[1] PHILIPS MED SYST,BEST,NETHERLANDS
关键词
interventional MR; catheter visualization; MR angiography;
D O I
10.1002/mrm.1910380214
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new technique for visualization of interventional devices using MR is presented, A prototype catheter was equipped with a thin copper wire loop, leading from the proximal end to the tip and back, A small current (10-150 mA) through these two parts of a wire induces a local magnetic field along the catheter, Introduction of this catheter into the main magnetic field of the MR imager locally disturbs the homogeneity of the magnetic field, Image locations within the locally induced fields appear dark due to signal loss, and the extent of this effect can be varied during the procedure by simply adjusting the current, Different dedicated wire configurations allow visualization of the catheter in its whole length or in parts, i.e., with markers for balloons, Fast gradient echo sequences that provide a bright signal from inflowing blood are used for rapid imaging.
引用
收藏
页码:253 / 258
页数:6
相关论文
共 13 条
[1]  
ACKERMAN L, 1986, P SMRM 5 ANN M MONTR, P1131
[2]  
BAKKER JG, 1996, P ISMRM 14 ANN M NEW, P900
[3]  
DARKAZANLI A, 1992, P SMRM 11 ANN M BERL, P591
[4]   REAL-TIME POSITION MONITORING OF INVASIVE DEVICES USING MAGNETIC-RESONANCE [J].
DUMOULIN, CL ;
SOUZA, SP ;
DARROW, RD .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (03) :411-415
[5]   MRI-MONITORED CRYOSURGERY IN THE RABBIT BRAIN [J].
GILBERT, JC ;
RUBINSKY, B ;
ROOS, MS ;
WONG, STS ;
BRENNAN, KM .
MAGNETIC RESONANCE IMAGING, 1993, 11 (08) :1155-1164
[6]   MR IMAGING OF LASER-TISSUE INTERACTIONS [J].
JOLESZ, FA ;
BLEIER, AR ;
JAKAB, P ;
RUENZEL, PW ;
HUTTL, K ;
JAKO, GJ .
RADIOLOGY, 1988, 168 (01) :249-253
[7]   VASCULAR INTERVENTIONS GUIDED BY ULTRAFAST MR-IMAGING - EVALUATION OF DIFFERENT MATERIALS [J].
KOCHLI, VD ;
MCKINNON, GC ;
HOFMANN, E ;
VONSCHULTHESS, GK .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (03) :309-314
[8]  
LENZ G, 1996, P ISMRM 4 SCI M EXH, P901
[9]   INTRAVASCULAR MR TRACKING CATHETER - PRELIMINARY EXPERIMENTAL EVALUATION [J].
LEUNG, DA ;
DEBATIN, JF ;
WILDERMUTH, S ;
MCKINNON, GC ;
HOLTZ, D ;
DUMOULIN, CL ;
DARROW, RD ;
HOFMANN, E ;
VONSCHULTHESS, GK .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 164 (05) :1265-1270
[10]   REAL-TIME BIPLANAR NEEDLE TRACKING FOR INTERVENTIONAL MR-IMAGING PROCEDURES [J].
LEUNG, DA ;
DEBATIN, JF ;
WILDERMUTH, S ;
HESKE, N ;
DUMOULIN, CL ;
DARROW, RD ;
HAUSER, M ;
DAVIS, CP ;
VONSCHULTHESS, GK .
RADIOLOGY, 1995, 197 (02) :485-488