Removal of local field gradient artefacts in BOLD contrast imaging of head and neck tumours

被引:8
作者
Demeure, RJ
Jordan, BF
Yang, QX
Beghein, N
Smith, MB
Grégoire, V
Gallez, B
机构
[1] Univ Catholique Louvain, Lab Biomed Magnet Resonance, B-1200 Brussels, Belgium
[2] Univ Catholique Louvain, Lab Med Chem & Radiopharm, B-1200 Brussels, Belgium
[3] Penn State Univ, Coll Med, Dept Radiol, Ctr NMR Res,Hershey Med Ctr, Hershey, PA 17033 USA
[4] St Luc Univ Hosp, Dept Radiat Oncol, B-1200 Brussels, Belgium
[5] Univ Catholique Louvain, Lab Radiobiol & Radioprot, B-1200 Brussels, Belgium
关键词
D O I
10.1088/0031-9155/47/10/315
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Monitoring of oxygenation in tumours is an important issue in predicting the success of anti-cancer treatments such as radiotherapy. Gradient echo (GE) imaging sequences can be used for monitoring changes in tumour blood flow and oxygenation. However. the application of this method in head and neck tumours is hampered by significant artefacts and losses of the MR signal near air-tissue interfaces. We investigated the usefulness of a gradient-echo slice excitation profile (GESEPI) sequence that should keep the oxygen contrast while recovering the signal loss caused by susceptibility artefacts. A tumour model was implanted in the neck and in the leg of mice. MR imaging was performed at 4.7 T. GE and GESEPI sequences were used for monitoring the blood oxygen level dependent (BOLD) contrast after carbogen breathing. The pO(2) was also monitored in tumours using an OxyLite probe (Oxford Optronics). Using the tumours implanted in the leg. A e found that the variations of signal intensity after carbogen breathing were similar in both sequences. In the tumour implanted in the neck, it was possible. using GESEPI sequences. to recover the signal loss caused by susceptibility artefacts and to monitor the effect of carbogen-induced changes in the tumour.
引用
收藏
页码:1819 / 1825
页数:7
相关论文
共 19 条
[11]  
Robinson SP, 1999, NMR BIOMED, V12, P98, DOI 10.1002/(SICI)1099-1492(199904)12:2<98::AID-NBM556>3.0.CO
[12]  
2-I
[13]   NONINVASIVE MONITORING OF CARBOGEN-INDUCED CHANGES IN TUMOR BLOOD-FLOW AND OXYGENATION BY FUNCTIONAL MAGNETIC-RESONANCE-IMAGING [J].
ROBINSON, SP ;
HOWE, FA ;
GRIFFITHS, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (04) :855-859
[14]   THE HISTOLOGICAL STRUCTURE OF SOME HUMAN LUNG CANCERS AND THE POSSIBLE IMPLICATIONS FOR RADIOTHERAPY [J].
THOMLINSON, RH ;
GRAY, LH .
BRITISH JOURNAL OF CANCER, 1955, 9 (04) :539-&
[15]   ECHO-PLANAR TIME COURSE MRI OF CAT BRAIN OXYGENATION CHANGES [J].
TURNER, R ;
LEBIHAN, D ;
MOONEN, CTW ;
DESPRES, D ;
FRANK, J .
MAGNETIC RESONANCE IN MEDICINE, 1991, 22 (01) :159-166
[16]  
van der Sanden BPJ, 1999, MAGNET RESON MED, V42, P490
[17]   Removal of local field gradient artifacts in T2*-weighted images at high fields by gradient-echo slice excitation profile imaging [J].
Yang, QX ;
Williams, GD ;
Demeure, RJ ;
Mosher, TJ ;
Smith, MB .
MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (03) :402-409
[18]  
Yang QX, 1999, MAGNET RESON MED, V41, P423, DOI 10.1002/(SICI)1522-2594(199902)41:2<423::AID-MRM30>3.0.CO
[19]  
2-M