ECAT ART - A continuously rotating PET camera: Performance characteristics, initial clinical studies, and installation considerations in a nuclear medicine department

被引:74
作者
Bailey, DL
Young, H
Bloomfield, PM
Meikle, SR
Glass, D
Myers, MJ
Spinks, TJ
Watson, CC
Luk, P
Peters, AM
Jones, T
机构
[1] HAMMERSMITH HOSP, DEPT NUCL MED, LONDON, ENGLAND
[2] CTI PET SYST, KNOXVILLE, TN USA
关键词
three-dimensional image reconstruction; partial-ring scanner; position emission tomography;
D O I
10.1007/BF01728302
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Advances in fully three-dimensional (3D) image reconstruction techniques have permitted the development of a commercial, rotating, partial ring, fully 3D positron emission tomographic (PET) scanner, the ECAT ART. The system has less than one-half the number of bismuth germanate detectors compared with a full ring scanner with the equivalent field of view, resulting in reduced capital cost. The performance characteristics, implications for installation in a nuclear medicine department, and clinical utility of the scanner are presented in this report. The sensitivity (20 cm diameterx20 cm long cylindrical phantom, no scatter correction) is 11400 cps . kBq(-1). ml(-1). This compares with 5800 and 40500 cps . kBq(-1). ml(-1) in 2D and 3D respectively for the equivalent full ring scanner (ECAT EXACT). With an energy window of 350-650 keV the maximum noise equivalent count (NEC) rate was 27 kcps at a radioactivity concentration of similar to 15 kBq . ml(-1) in the cylinder. Spatial resolution is similar to 6 mm full width at half maximum on axis degrading to just under 8 mm at a distance of 20 cm off axis. Installation and use within the nuclear medicine department does not appreciably increase background levels of radiation on gamma cameras in adjacent rooms and the dose rate to an operator in the same room is 2 mu Sv . h(-1) for a typical fluorine-18 fluorodeoxyglucose (F-18-FDG) study with an initial injected activity of 370 MBq. The scanner has been used for clinical imaging with F-18-FDG for neurological and oncological applications. Its novel use for imaging iron-52 transferrin for localising erythropoietic activity demonstrates its sensitivity and resolution advantages over a conventional dual-headed gamma camera. The ECAT ART provides a viable alternative to conventional full ring PET scanners without compromising the performance required for clinical PET imaging.
引用
收藏
页码:6 / 15
页数:10
相关论文
共 37 条
[1]   A METHOD FOR MEASURING THE ABSOLUTE SENSITIVITY OF POSITRON EMISSION TOMOGRAPHIC SCANNERS [J].
BAILEY, DL ;
JONES, T ;
SPINKS, TJ .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1991, 18 (06) :374-379
[2]   NOISE EQUIVALENT COUNT MEASUREMENTS IN A NEURO-PET SCANNER WITH RETRACTABLE SEPTA [J].
BAILEY, DL ;
JONES, T ;
SPINKS, TJ ;
GILARDI, MC ;
TOWNSEND, DW .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) :256-260
[3]  
BAILEY DL, 1993, INT CONGR SER, V1030, P341
[4]  
BAILEY DL, 1993, J NUCL MED, V34
[5]  
Carroll LR, 1983, EMISSION COMPUTED TO, P235
[6]  
CARSON RE, 1995, J NUCL MED, V36
[7]   3D PET USING A CONVENTIONAL MULTISLICE TOMOGRAPH WITHOUT SEPTA [J].
CHERRY, SR ;
DAHLBOM, M ;
HOFFMAN, EJ .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (04) :655-668
[8]   IMPROVED DETECTION OF FOCAL CEREBRAL BLOOD-FLOW CHANGES USING 3-DIMENSIONAL POSITRON EMISSION TOMOGRAPHY [J].
CHERRY, SR ;
WOODS, RP ;
HOFFMAN, EJ ;
MAZZIOTTA, JC .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1993, 13 (04) :630-638
[9]  
CHESLER DA, 1971, J NUCL MED, V12, P347
[10]   FULLY 3-DIMENSIONAL POSITRON EMISSION TOMOGRAPHY [J].
COLSHER, JG .
PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (01) :103-115