A modified Z-type flow-through cell for optical, electrochemical, and optoelectrochemical flow injection analysis measurements

被引:1
作者
Haghighi, B [1 ]
Maleki, N [1 ]
Safavi, A [1 ]
机构
[1] SHIRAZ UNIV,COLL SCI,DEPT CHEM,SHIRAZ 71454,IRAN
关键词
D O I
10.1006/mchj.1997.1484
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The design and construction of a modified Z-type flow-through cell that can be used for optical, electrochemical, and optoelectrochemical measurements are reported. In optical analysis, the proposed flow cell is placed in the light path of a spectrophotometer and worked as a commercial Z-type optical Row cell. In electrochemical analysis, the working and auxiliary electrodes are two platinum sheets placed at both ends of the cell cavity; the cell cavity is enclosed by two glass windows. The reference electrode is a saturated calomel electrode placed downstream in an overflow tube. The performance of the proposed electrochemical Row cell was studied by analysis of hexacyanoferrate solution as an electroactive species. In optoelectrochemical analysis, the proposed flow cell is operated optically and electrochemically, simultaneously. The flow cell is placed in the light path of a spectrophotometer, constant potential is applied to the electrodes, and the optical and electrochemical signals are recorded simultaneously by the use of two recorders. (C) 1997 Academic Press.
引用
收藏
页码:339 / 345
页数:7
相关论文
共 10 条
[1]  
*BIOAN SYST INC, 1977, APPL NOT LIQ CHROM E
[2]   STUDY OF A GLASSY-CARBON ELECTRODE IN AMPEROMETRIC DETECTION USING DC, NORMAL AND DIFFERENTIAL PULSE TECHNIQUES [J].
DIEKER, JW ;
VANDERLINDEN, WE ;
POPPE, H .
TALANTA, 1979, 26 (07) :511-518
[3]  
HEINEMAN WR, 1984, LAB TECHNIQUES ELECT
[4]   DETERMINATION OF READILY OXIDIZED COMPOUNDS BY FLOW INJECTION ANALYSIS AND REDOX POTENTIAL DETECTION [J].
KARLBERG, B ;
THELANDER, S .
ANALYST, 1978, 103 (1232) :1154-1159
[5]  
KOIFE RC, 1979, ANAL CHEM, V61, P741
[6]   AMPEROMETRIC FLOW-THROUGH WIRE DETECTOR - A PRACTICAL DESIGN WITH HIGH-SENSITIVITY [J].
LOWN, JA ;
KOILE, R ;
JOHNSON, DC .
ANALYTICA CHIMICA ACTA, 1980, 116 (01) :33-39
[7]   Design of a new phase separator for liquid-liquid extraction in flowing systems [J].
Maleki, N ;
Haghighi, B ;
Safavi, A .
MICROCHEMICAL JOURNAL, 1996, 53 (02) :147-151
[8]   MERCURY-BASED ELECTROCHEMICAL DETECTOR FOR LIQUID-CHROMATOGRAPHY FOR DETECTION OF GLUTATHIONE AND OTHER SULFUR-CONTAINING-COMPOUNDS [J].
RABENSTEIN, DL ;
SAETRE, R .
ANALYTICAL CHEMISTRY, 1977, 49 (07) :1036-1039
[9]  
Ruzicka Jaromir., 1988, FLOW INJECTION ANAL, V2nd
[10]  
STROBEL HA, 1989, CHEM INSTRUMENTATION