Optimization of beam filtering, kV-mA regulation curve, and image intensifier entrance exposure rate to reduce radiation exposure in angiographic fluoroscopy.

被引:4
作者
Barkhausen, J
Schoenfelder, D
Nagel, HD
Stöblen, F
Müller, RD
机构
[1] Univ Essen Gesamthsch Klinikum, Zent Inst Rontgendiagnost, D-45122 Essen, Germany
[2] Philips Med Syst, Hamburg, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 1999年 / 171卷 / 05期
关键词
fluoroscopy; digital subtraction angiography; dose reduction;
D O I
10.1055/s-1999-261
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aim of study: Evaluation of radiation exposure and image quality during fluoroscopy using a new vascular X-ray system. Material and methods: The measurements were made on an Integris V 3000 X-ray system with MRC tube and SpectraBeam technology (Philips Medical Systems, Hamburg). Entrance dose rates were measured with phantoms for the three fluoroscopy levels (1-3) which differed with regard to beam filtering and image intensiver entrance exposure rate. We evaluated 132 diagnostic and interventional angiographic studies. The angiographic investigators were asked to start with level 1 and to change to the next fluoroscopy level only in the case of insufficient image quality. Results: Entrance dose rate is reduced by approx. 74% at fluoroscopy level 1 and by approx. 46% at level Z relative to level 3 which is comparable to angiographic X-ray systems without MRC tube and SpectraBeam technology. Because level 1 ensured a sufficient image quality in 92% of the diagnostic and 60% of the interventional angiographic procedures a change to higher fluoroscopy levels was not necessary. Conclusion: Reduction of the intensifier exposure rate and the optimization of beam filtering enabled us to reduce the radiation exposure considerably. The procedure was well accepted by the angiographic investigators due to the diagnostically sufficient image quality of the fluoroscopy level 1.
引用
收藏
页码:391 / 395
页数:5
相关论文
共 15 条
[1]  
AMMAN E, 1997, BRIT J RADIOL, V70, P1
[2]   REDUCTION OF RADIATION EXPOSURE WHILE MAINTAINING HIGH-QUALITY FLUOROSCOPIC IMAGES DURING INTERVENTIONAL CARDIOLOGY USING NOVEL X-RAY TUBE TECHNOLOGY WITH EXTRA BEAM FILTERING [J].
DENBOER, A ;
DEFEYTER, PJ ;
HUMMEL, WA ;
KEANE, D ;
ROELANDT, JRTC .
CIRCULATION, 1994, 89 (06) :2710-2714
[3]  
EWEN K, 1994, Z KARDIOL, V83, P57
[4]   Reduction of radiation dose in pediatric patients using pulsed fluoroscopy [J].
Hernandez, RJ ;
Goodsitt, MM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 167 (05) :1247-1253
[5]   Patient dose measurements from femoral angiography [J].
Hoskins, PR ;
Gillespie, I ;
Ireland, HM .
BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (828) :1159-1164
[6]  
Imhof K, 1995, Aktuelle Radiol, V5, P385
[7]  
Katsuda T, 1996, Radiol Technol, V68, P18
[8]   Effective doses to members of the public from the diagnostic application of ionizing radiation in Germany [J].
Kaul, A ;
Bauer, B ;
Bernhardt, J ;
Nosske, D ;
Veit, R .
EUROPEAN RADIOLOGY, 1997, 7 (07) :1127-1132
[9]   Radiodermatitis after transjugular intrahepatic portosystemic shunt [J].
Knautz, MA ;
Abele, DC ;
Reynolds, TL .
SOUTHERN MEDICAL JOURNAL, 1997, 90 (03) :352-356
[10]   Reduction of radiation exposure for patient and examiner in interventional angiography. [J].
Pecher, G ;
Koenig, H ;
Pecher, S ;
Gosch, D ;
Voigt, P ;
Schulz, HG .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1998, 169 (05) :505-509