Reduction of radiation exposure for patient and examiner in interventional angiography.

被引:8
作者
Pecher, G
Koenig, H
Pecher, S
Gosch, D
Voigt, P
Schulz, HG
机构
[1] Univ Leipzig, Klin & Poliklin Diagnost Radiol, D-04103 Leipzig, Germany
[2] Stadt Klinikum Dessau, Inst Radiol, Dessau, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 1998年 / 169卷 / 05期
关键词
X-ray dose; interventional vascular procedures; pulsed fluoroscopy; last-image hold; road mapping mode;
D O I
10.1055/s-2007-1015328
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In a retrospective analysis of vascular interventional procedures, relations between parameters of the examination and radiation exposure of patient and medical personnel are examined. Material and Method: 1208 vascular interventional procedures are evaluated. Interventional procedures are divided into three groups: percutaneous transluminal angioplasty, implantation of a stent, thrombolysis. Results: Mean values;of the radiation dose of patient and radiology personnel are reported for these examinations. The mean Value of the radiation dose of the physician was 7 mu Sv (maximum 24 mu Sv), that of the patient 1548 cGy . cm(2) (maximum 8485 cGy . cm(2)). Conclusion: The quantity of X-rays to the patient may be lowered by using pulsed fluoroscopy and by reducing the number of radiographs. Reduction of the number of radiographs may be achieved by using the last-image hold and the road mapping mode. The operator's dose can be decreased by using additional radiation protection systems like a MAVIG(R)-radiation protection wall. The radiation dose reduction was 61% for the physician and 17% for the patient.
引用
收藏
页码:505 / 509
页数:5
相关论文
共 22 条
[1]   PERCEPTUAL COMPARISON OF PULSED AND CONTINUOUS FLUOROSCOPY [J].
AUFRICHTIG, R ;
XUE, P ;
THOMAS, CW ;
GILMORE, GC ;
WILSON, DL .
MEDICAL PHYSICS, 1994, 21 (02) :245-256
[2]  
*BUND UMW, 1997, BUND BEK EMPF STRAHL
[3]   Radiation exposure to the hands of operators during angiographic procedures [J].
Damilakis, J ;
Koukourakis, M ;
Hatjidakis, A ;
Karabekios, S ;
Gourtsoyiannis, N .
EUROPEAN JOURNAL OF RADIOLOGY, 1995, 21 (01) :72-75
[4]  
EDER H, 1995, RADIOLOGE, V35, P156
[5]  
EWEN K, 1994, Z KARDIOL, V83, P57
[6]   AN ASSESSMENT OF THE RADIATION-DOSE RECEIVED BY STAFF USING FLUOROSCOPIC EQUIPMENT [J].
FAULKNER, K ;
MOORES, BM .
BRITISH JOURNAL OF RADIOLOGY, 1982, 55 (652) :272-276
[7]   RADIATION EXPOSURE OF RADIOLOGISTS DURING ANGIOGRAPHY - DOSE MEASUREMENTS OUTSIDE THE LEAD APRON [J].
FISCHER, H ;
PRZETAK, C ;
TEUBERT, G ;
EWEN, K ;
MODDER, U .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1995, 162 (02) :152-156
[8]  
*GEM MIN, 1994, 7 GEM MIN, P286
[9]   First experiences with pulsed fluoroscopy using a multifunctional fluoroscopy unit. [J].
Herrmann, K ;
Helmberger, T ;
Waggershauser, T ;
Schatzl, M ;
Allmendinger, H ;
Reiser, M .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1996, 165 (05) :475-479
[10]  
HERRMANN K, 1994, ELEKTROMEDICA, V62, P56