A convenient approach for finite-element-analyses of orthopaedic implants in bone contact: Modeling and experimental validation

被引:99
作者
Kluess, Daniel [1 ]
Souffrant, Robert [1 ]
Mittelmeier, Wolfram [1 ]
Wree, Andreas [3 ]
Schmitz, Klaus-Peter [2 ]
Bader, Rainer [1 ]
机构
[1] Univ Rostock, Dept Orthopaed, D-18057 Rostock, Germany
[2] Univ Rostock, Inst Biomed Engn, D-18057 Rostock, Germany
[3] Univ Rostock, Dept Anat, D-18057 Rostock, Germany
关键词
Finite-element-analysis; Biomechanics; Orthopaedics; Bone; Implant; TOTAL KNEE ARTHROPLASTY; HIP-JOINT; MECHANICAL-PROPERTIES; LOAD-TRANSFER; PELVIC BONE; COMPONENT; BEHAVIOR; FORCES;
D O I
10.1016/j.cmpb.2009.01.004
中图分类号
TP39 [计算机的应用];
学科分类号
080201 [机械制造及其自动化];
摘要
With regard to the growing potential of finite-element-analysis (FEA) in the field of orthopaedic biomechanics, we present an approach helping in the development of appropriate models of the implant-bone compound. The algorithm is based on computed-tomography data of the bone and accordant computer-aided-design (CAD) data of the implant and aims at predicting the bone strains and interface mechanics of the included parts. The developed algorithm was validated exemplary using an acetabular cup in combination with a left and a right fresh-frozen human hemipelvis. The strains under maximum loads during the gait cycle as well as the micromotion in the bone-implant interface were measured and compared to results from equivalent finite-element-analyses. Thereby, we found strong correlation between the calculated and measured principal strains with correlation coefficients of r(2) = 0.94 (left side) and r(2) = 0.86 (right side). A validation of micromotion was not possible due to limited accuracy of the motion tracking system. (C) 2009 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:23 / 30
页数:8
相关论文
共 24 条
[1]
ABAQUS, 2007, VERS 6 7 DOC ABAQUS
[2]
Subject-specific finite element model of the pelvis: Development, validation and sensitivity studies [J].
Anderson, AE ;
Peters, CL ;
Tuttle, BD ;
Weiss, JA .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (03) :364-373
[3]
Finite element contact analysis of the hip joint [J].
Bachtar, Fuziansyah ;
Chen, Xian ;
Hisada, Toshiaki .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2006, 44 (08) :643-651
[4]
HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[5]
Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[6]
Tibial forces measured in vivo after total knee arthroplasty [J].
D'Lima, DD ;
Patil, S ;
Steklov, N ;
Slamin, JE ;
Colwell, CW .
JOURNAL OF ARTHROPLASTY, 2006, 21 (02) :255-262
[7]
MECHANICAL AND TEXTURAL PROPERTIES OF PELVIC TRABECULAR BONE [J].
DALSTRA, M ;
HUISKES, R ;
ODGAARD, A ;
VANERNING, L .
JOURNAL OF BIOMECHANICS, 1993, 26 (4-5) :523-535
[8]
DEVELOPMENT AND VALIDATION OF A 3-DIMENSIONAL FINITE-ELEMENT MODEL OF THE PELVIC BONE [J].
DALSTRA, M ;
HUISKES, R ;
VANERNING, L .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1995, 117 (03) :272-278
[9]
LOAD-TRANSFER ACROSS THE PELVIC BONE [J].
DALSTRA, M ;
HUISKES, R .
JOURNAL OF BIOMECHANICS, 1995, 28 (06) :715-724
[10]
Analysis of the structural behavior of the pelvis during lateral impact using the finite element method [J].
Dawson, JM ;
Khmelniker, BV ;
McAndrew, MP .
ACCIDENT ANALYSIS AND PREVENTION, 1999, 31 (1-2) :109-119