Single standard calibration and data processing in flow injection titration based on concentration gradients

被引:12
作者
Araujo, MCU [1 ]
Santos, AV [1 ]
Honorato, RS [1 ]
机构
[1] UNIV ESTADUAL CAMPINAS, INST QUIM, BR-13081970 CAMPINAS, SP, BRAZIL
来源
JOURNAL OF AUTOMATIC CHEMISTRY | 1997年 / 19卷 / 05期
关键词
D O I
10.1155/S1463924697000163
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This paper describes use of gradients of concentration generated in flow injection (FT) systems to perform determinations based on points where the concentration of titrant and analyte are at stoichiometric ratio. Two procedures were developed. In one procedure the titrant is injected in a FI manifold and merges with the sample which is continuously pumped towards the detector. In the other procedure the sample is injected and merged with the titrant which is continuously pumped. Both techniques make use of concentration gradients of the sample or titrant generated in FI manifolds that contain a mixing chamber. This gradient is calibrated employing only one standard solution (usually the titrant) in order to convert any detector signal, obtained in the elapsed time after injection, to instantaneous concentration values. The flow system is microcomputer controlled and data are treated to locate points where the concentration of titrant and analyte are at the stoichiometric ratio. These points are found in abrupt changes of the signal x concentration curves obtained in the presence of the reaction. The method has been evaluated for determination of Fe(II) and acetic acid by, spectrophotometric and conductimetric detection, respectively. Results shore, a mean relative standard deviation bower than 1%, an average accuracy of 1% and a high sampling processing capability (40 to 60 samples pet hour).
引用
收藏
页码:157 / 164
页数:8
相关论文
共 21 条
  • [11] GRADIENT TITRATION - NOVEL APPROACH TO CONTINUOUS MONITORING USING ION-SELECTIVE ELECTRODES
    FLEET, B
    HO, AYW
    [J]. ANALYTICAL CHEMISTRY, 1974, 46 (01) : 9 - 11
  • [12] FLOW-INJECTION ANALYSIS BY USING ACID-BASE OR METAL-LIGAND BUFFERS AND COLOR-INDICATORS
    ISHIBASHI, N
    IMATO, T
    [J]. FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1986, 323 (03): : 244 - 248
  • [13] BINARY SEARCH IN FLOW TITRATION EMPLOYING PHOTOMETRIC END-POINT DETECTION
    KORN, M
    GOUVEIA, LFBP
    DEOLIVEIRA, E
    REIS, BF
    [J]. ANALYTICA CHIMICA ACTA, 1995, 313 (03) : 177 - 184
  • [14] Malcolme-Lawes D. J., 1989, Laboratory Microcomputer, V8, P44
  • [15] NAGY G, 1966, ANAL CHIM ACTA, V91, P87
  • [16] Pasquini C., 1990, Laboratory Microcomputer, V9, P44
  • [17] OPERATOR-FREE FLOW-INJECTION ANALYZER
    PASQUINI, C
    DEFARIA, LC
    [J]. JOURNAL OF AUTOMATIC CHEMISTRY, 1991, 13 (04): : 143 - 146
  • [18] FLOW INJECTION ANALYSIS .9. NEW APPROACH TO CONTINUOUS-FLOW TITRATIONS
    RUZICKA, J
    HANSEN, EH
    MOSBAEK, H
    [J]. ANALYTICA CHIMICA ACTA, 1977, 92 (SEP) : 235 - 249
  • [19] Ruzicka Jaromir., 1988, FLOW INJECTION ANAL, V2nd
  • [20] Standard additions in flow injection analysis based on merging zones and gradient exploitation: Application to copper determination in spirits
    Silva, EC
    Araujo, MCU
    Honorato, RS
    Lima, JLFC
    Zagatto, EAG
    Brienza, SMB
    [J]. ANALYTICA CHIMICA ACTA, 1996, 319 (1-2) : 153 - 158