Mechanical tests and finite element models for bone holding power of tibial locking screws

被引:39
作者
Hou, SM
Hsu, CC
Wang, JL
Chao, CK
Lin, J [1 ]
机构
[1] Natl Taiwan Univ Hosp, Dept Orthoped Surg, Taipei, Taiwan
[2] Natl Taiwan Univ Sci & Technol, Dept Mech Engn, Taipei, Taiwan
关键词
finite element analysis; holding power; tibial screw;
D O I
10.1016/j.clinbiomech.2004.04.012
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Objective. To investigate the bone holding power of tibial locking screws. Design. The bone holding power was assessed by mechanical testing and finite element analysis. Background. Screw loosening might threaten fracture fixation and bone healing. Methods. In mechanical tests, six types of different tibial locking screws were inserted into low-density polyurethane foam tubes, which simulated osteoporotic bone. The screws were Pushed out of the foam bone by an axial load, and the maximal pushout load was recorded. In finite element analysis, three-dimensional finite element models with a nonlinear contact interface between the screws and the bones were created to simulate the mechanical testing. The total strain energy of the bone and total reaction force of the screws were recorded. The contribution of the design factors was analyzed by the Taguchi method. Results. In the mechanical tests, foam bone was stripped by the screw threads without screw deformation. The testing results were closely related to those of finite element analysis. The Taguchi analysis showed that the descending order of contribution of the design factors was outer diameter, pitch, half angle, and inner diameter. Root radius and thread width had minimal effects. Conclusions. The bone holding power of the screws could be reliably assessed by finite element models, which could analyze the effects of all the design factors independently and were potentially applicable to screws with irregular thread patterns. Relevance The finite element models built in this study may help manufacturers in evaluating new designs of locking screws and assist surgeons in selecting suitable devices for patients with severe osteoporosis. (C) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:738 / 745
页数:8
相关论文
共 21 条
[1]
[Anonymous], 2003, ANSYS 7 0 DOC
[2]
Cancellous bone screw thread design and holding power [J].
Asnis, SE ;
Ernberg, JJ ;
Bostrom, MPG ;
Wright, TM ;
Harrington, RM ;
Tencer, A ;
Peterson, M .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1996, 10 (07) :462-469
[3]
Mechanical performance of standard and cannulated 4.0-mm cancellous bone screws [J].
Brown, GA ;
McCarthy, T ;
Bourgeault, CA ;
Callahan, DJ .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2000, 18 (02) :307-312
[4]
Brumback RJ, 1996, CLIN ORTHOP RELAT R, P292
[5]
Factors affecting the pullout strength of cancellous bone screws [J].
Chapman, JR ;
Harrington, RM ;
Lee, KM ;
Anderson, PA ;
Tencer, AF ;
Kowalski, D .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (03) :391-398
[6]
Statistical methods in finite element analysis [J].
Dar, FH ;
Meakin, JR ;
Aspden, RM .
JOURNAL OF BIOMECHANICS, 2002, 35 (09) :1155-1161
[7]
DeCoster T A, 1990, J Orthop Trauma, V4, P169, DOI 10.1097/00005131-199004020-00012
[8]
FOWLKES WY, 1995, ENG METHODS ROBUST P, P292
[9]
HALSEY D, 1992, CLIN ORTHOP RELAT R, P305
[10]
HAYES WC, 1991, BASIC ORTHOPAEDIC BI, P93