Step activity monitor: Increased accuracy in quantifying ambulatory activity

被引:135
作者
Shepherd, EF
Toloza, E
McClung, CD
Schmalzried, TP
机构
[1] Orthoped Hosp, Joint Replacement Inst, Los Angeles, CA 90007 USA
[2] Univ Calif Los Angeles, Harbor Med Ctr, Dept Orthopaed Surg, Torrance, CA 90509 USA
关键词
D O I
10.1002/jor.1100170512
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This study compares the accuracy of a two-dimensional accelerometer worn on the ankle (a step activity monitor) with that of an electronic, digital pedomrter worn on the belt line. Twenty-nine human subjects were evaluated while they briskly walked 400 M, slowly walked 10 M, and ascended and descended a flight of stairs. The step activity monitor had less error in all activities: its mean absolute error was 0.54%, whereas that of the pedometer was 2.82%. The difference was more pronounced in obese subjects (body mass index greater than 30), with an overall mean absolute error of 0.48% for the step activity monitor and 6.12% for the pedometer (nearly 13 times that of the step activity monitor). For subjects with a body mass index less than 30, the step activity monitor had an overall error of 0.56% and the pedometer had an overall error of 1.56% (less than 3 times that of the step activity monitor). The absolute error of the pedometer was positively correlated with body mass index (r = 0.792, p < 0.0001) and weight (r = 0.753, p < 0.0001), whereas the error of the step activity monitor was not significantly correlated with either. Neither device was significantly biased by age, gender, or the presence of a lower-extremity joint prosthesis. The accuracy and additional capabilities, including a real-time memory record of activity. of the step activity monitor make it well suited for objectively quantifying ambulatory activity, especially for obese subjects.
引用
收藏
页码:703 / 708
页数:6
相关论文
共 33 条