Quantitative Ex Vivo and In Vivo Comparison of Lumen Dimensions Measured by Optical Coherence Tomography and Intravascular Ultrasound in Human Coronary Arteries

被引:92
作者
Gonzalo, Nieves [1 ]
Serruys, Patrick W. [1 ]
Garcia-Garcia, Hector M. [1 ]
van Soest, Gijs [1 ]
Okamura, Takayuki [1 ]
Ligthart, Jurgen [1 ]
Knaapen, Michiel [3 ]
Verheye, Stefan [2 ]
Bruining, Nico [1 ]
Regar, Evelyn [1 ]
机构
[1] Erasmus MC, Thoraxctr, Dept Cardiol, NL-3015 CE Rotterdam, Netherlands
[2] Middelheim Hosp, Antwerp Cardiovasc Inst, Antwerp, Belgium
[3] Histogenex, Antwerp, Belgium
来源
REVISTA ESPANOLA DE CARDIOLOGIA | 2009年 / 62卷 / 06期
关键词
Optical coherence tomography; Intravascular ultrasound; Lumen dimensions; Ex vivo; In vivo; Intracoronary imaging; SIROLIMUS-ELUTING STENT; TAXUS-II TRIAL; NEOINTIMAL COVERAGE; IMPLANTATION; TERM; CULPRIT; IMAGES;
D O I
10.1016/S0300-8932(09)71328-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction and objectives. The relationship between the lumen dimensions obtained in human coronary arteries using intravascular ultrasound (IVUS) and those obtained using optical coherence tomography (OCT) is not well understood. The objectives were to compare the lumen measurements obtained ex vivo in human coronary arteries using IVUS, OCT and histomorphometry, and in vivo in patients using IVUS and OCT with and without balloon occlusion. Methods. Ex vivo study: the lumen areas of matched anatomical sections of human coronary arteries were measured using IVUS, OCT and histology. In vivo study: the lumen areas in matched sections were measured using IVUS and OCT with and without occlusion. Results. Ex vivo: in the eight specimens studied, the lumen area obtained using OCT and IVUS was larger than that obtained using histomorphometry: mean difference 0.8+/-1 mm(2) (28%) for OCT and 1.3+/-1.1 mm(2) (40%) for IVUS. In vivo: in the five vessels analyzed, the lumen area obtained using IVUS was larger than that obtained using OCT: mean difference 1.67+/-0.54 mm(2) (33.7%) for IVUS relative to OCT with occlusion and 1.11+/-0.53 mm(2) (21.5%) relative to OCT without occlusion. The lumen area obtained using OCT without occlusion was larger than that obtained using OCT with occlusion: mean difference 0.61+/-0.23 mm(2) (13%). Conclusions. In fixed human coronary arteries, both IVUS and OCT overestimated the lumen area compared with histomorphometry. In vivo the lumen dimensions obtained using IVUS were larger than those obtained using OCT, with or without occlusion. Moreover, the OCT image acquisition technique (i.e. with or without occlusion) also had an impact on lumen measurement.
引用
收藏
页码:615 / 624
页数:10
相关论文
共 29 条
[11]   In vivo volumetric analysis of coronary stent using optical coherence tomography with a novel balloon occlusion-flushing catheter:: A comparison with intravascular ultrasound [J].
Kawase, Y ;
Hoshino, K ;
Yoneyama, R ;
McGregor, J ;
Hajjar, RJ ;
Jang, IK ;
Hayase, M .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2005, 31 (10) :1343-1349
[12]   Assessment of culprit lesion morphology in acute myocardial infarction - Ability of optical coherence tomography compared with intravascular ultrasound and coronary angioscopy [J].
Kubo, Takashi ;
Imanishi, Toshio ;
Takarada, Shigeho ;
Kuroi, Akio ;
Ueno, Satoshi ;
Yamano, Takashi ;
Tanimoto, Takashi ;
Matsuo, Yoshiki ;
Masho, Takashi ;
Kitabata, Hironori ;
Tsuda, Kazushi ;
Tomobuchi, Yoshiaki ;
Akasaka, Takashi .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 50 (10) :933-939
[13]   Assessment of coronary intima-media thickness by optical coherence tomography - Comparison with intravascular ultrasound [J].
Kume, T ;
Akasaka, T ;
Kawamoto, T ;
Watanabe, N ;
Toyota, E ;
Neishi, Y ;
Sukmawan, R ;
Sadahira, Y ;
Yoshida, K .
CIRCULATION JOURNAL, 2005, 69 (08) :903-907
[14]   Measurement of the thickness of the fibrous cap by optical coherence tomography [J].
Kume, Teruyoshi ;
Akasaka, Takashi ;
Kawamoto, Takahiro ;
Okura, Hiroyuki ;
Watanabe, Nozomi ;
Toyota, Eiji ;
Neishi, Yoji ;
Sukmawan, Renan ;
Sadahira, Yoshito ;
Yoshida, Kiyoshi .
AMERICAN HEART JOURNAL, 2006, 152 (04) :755.e1-755.e4
[15]   American College of Cardiology Clinical Expert Consensus Document on Standards for Acquisition, Measurement and Reporting of Intravascular Ultrasound Studies (IVUS) - A report of the American College of Cardiology Task Force on Clinical Expert Consensus - Documents developed in collaboration with the European Society of Cardiology endorsed by the Society of Cardiac Angiography and Interventions [J].
Mintz, GS ;
Nissen, SE ;
Anderson, WD ;
Bailey, SR ;
Erbel, R ;
Fitzgerald, PJ ;
Pinto, FJ ;
Rosenfield, K ;
Siegel, RJ ;
Tuzcu, EM ;
Yock, PG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (05) :1478-1492
[16]  
Prati Francesco, 2007, EuroIntervention, V3, P365, DOI 10.4244/EIJV3I3A66
[17]   From bench to bedside - A novel technique of acquiring OCT images [J].
Prati, Francesco ;
Cera, Maria ;
Ramazzotti, Vito ;
Imola, Fabrizio ;
Giudice, Rocco ;
Giudice, Marco ;
De Propris, Silvia ;
Albertucci, Mario .
CIRCULATION JOURNAL, 2008, 72 (05) :839-843
[18]  
REGAR E, 2009, INFORM HEAL IN PRESS
[19]   ASSESSMENT OF SHORT-TERM, MEDIUM-TERM, AND LONG-TERM VARIATIONS IN ARTERIAL DIMENSIONS FROM COMPUTER-ASSISTED QUANTITATION OF CORONARY CINEANGIOGRAMS [J].
REIBER, JHC ;
SERRUYS, PW ;
KOOIJMAN, CJ ;
WIJNS, W ;
SLAGER, CJ ;
GERBRANDS, JJ ;
SCHUURBIERS, JCH ;
DENBOER, A ;
HUGENHOLTZ, PG .
CIRCULATION, 1985, 71 (02) :280-288
[20]   Vascular responses at proximal and distal edges of paclitaxel-eluting stents - Serial intravascular ultrasound analysis from the TAXUS II trial [J].
Serruys, PW ;
Degertekin, M ;
Tanabe, K ;
Russell, ME ;
Guagliumi, G ;
Webb, J ;
Hamburger, J ;
Rutsch, W ;
Kaiser, C ;
Whitbourn, R ;
Camenzind, E ;
Meredith, I ;
Reeves, F ;
Nienaber, C ;
Benit, E ;
Disco, C ;
Koglin, J ;
Colombo, A .
CIRCULATION, 2004, 109 (05) :627-633