Evaluating x-ray detectors for radiographic applications: A comparison of ZnSCdS:Ag with Gd2O2S:Tb and Y2O2S:Tb screens

被引:89
作者
Kandarakis, I
Cavouras, D
Panayiotakis, GS
Nomicos, CD
机构
[1] TECHNOL EDUC INST ATHENS,DEPT MED INSTRUMENTAT TECHNOL,AIGALEO 12210,ATHENS,GREECE
[2] UNIV PATRAS,SCH MED,DEPT MED PHYS,PATRAS 26500,GREECE
[3] TECHNOL EDUC INST ATHENS,DEPT ELECT,ATHENS,GREECE
关键词
D O I
10.1088/0031-9155/42/7/009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
ZnSCdS:Ag was evaluated as a radiographic image receptor and was compared with Gd2O2:Tb and Y2O2S:Tb phosphors often used in radiography. The evaluation of a radiographic receptor was modelled as a three-step process: (i) determination of light output intensity as related to the input radiation dose, (ii) determination of visible light characteristics with respect to radiographic optical detectors, and (iii) determination of image information transfer efficiency. The light intensity emitted per unit of x-ray exposure rate was measured and theoretically calculated for laboratory prepared screens with coating thicknesses from 20 to 220 mg cm(-2) and tube voltages from 50 to 250 kVp. ZnSCdS:Ag light intensity was higher than that of Gd2O2S:Tb or Y2O2S:Tb for tube voltages less than 70 and 80 kVp respectively. ZnSCdS:Ag displayed the highest x-ray to light conversion efficiency (0.207) and had optical properties close to those of Gd2O2S:Tb and Y2O2S:Tb, and its emission spectrum was well matched to optical detectors. The image information transfer properties described by the modulation transfer function, the quantum noise transfer function, and the detective quantum efficiency were calculated for both general radiographic and mammographic conditions and were found to be intermediate between those of Gd2O2S:Tb and Y2O2S:Tb screens. Conclusively, ZnSCdS:Ag is an efficient phosphor well suited to radiography.
引用
收藏
页码:1351 / 1373
页数:23
相关论文
共 33 条
[1]  
[Anonymous], PHYS MED IMAGING REC
[2]   THE ROLE OF SCREEN PARAMETERS AND PRINT-THROUGH IN THE PERFORMANCE OF FILM-SCREEN SYSTEMS [J].
BEUTEL, J ;
APPLE, BA ;
SHAW, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (09) :1181-1193
[3]   ANALYSIS OF THE DETECTIVE QUANTUM EFFICIENCY OF A RADIOGRAPHIC SCREEN FILM COMBINATION [J].
BUNCH, PC ;
HUFF, KE ;
VANMETTER, R .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (05) :902-909
[4]   An evaluation of the Y2O3:Eu3+ scintillator for application in medical x-ray detectors and image receptors [J].
Cavouras, D ;
Kandarakis, I ;
Panayiotakis, GS ;
Evangelou, EK ;
Nomicos, CD .
MEDICAL PHYSICS, 1996, 23 (12) :1965-1975
[5]   ENERGY AND ANGULAR-DEPENDENCE OF X-RAY ABSORPTION AND ITS EFFECT ON RADIOGRAPHIC RESPONSE IN SCREEN-FILM SYSTEMS [J].
CHAN, HP ;
DOI, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (05) :565-579
[6]  
DAINTY JC, 1974, IMAGE SCI, P152
[7]   IMAGE QUALITY INDEX (IQI) FOR SCREEN FILM MAMMOGRAPHY [J].
DESPONDS, L ;
DEPEURSINGE, C ;
GRECESCU, M ;
HESSLER, C ;
SAMIRI, A ;
VALLEY, JF .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (01) :19-33
[8]   IMAGE-INFORMATION TRANSFER PROPERTIES OF X-RAY FLUORESCENT SCREENS [J].
DICK, CE ;
MOTZ, JW .
MEDICAL PHYSICS, 1981, 8 (03) :337-346
[9]   X-RAY-IMAGING WITH AMORPHOUS SELENIUM - DETECTIVE QUANTUM EFFICIENCY OF PHOTOCONDUCTIVE RECEPTORS FOR DIGITAL MAMMOGRAPHY [J].
FAHRIG, R ;
ROWLANDS, JA ;
YAFFE, MJ .
MEDICAL PHYSICS, 1995, 22 (02) :153-160
[10]  
GIAKOUMAKIS GE, 1991, APPL PHYS A-MATER, V52, P7